FIELD-BASED IDENTIFICATION OF MINERALS USING A BATTERY-OPERATED ELECTROCHEMICAL MEASURING SYSTEM WITH MECHANICAL TRANSFER OF THE SOLID TO A GRAPHITE ELECTRODE

被引:21
作者
BOND, AM [1 ]
SCHOLZ, F [1 ]
机构
[1] DEAKIN UNIV,DEPT CHEM & ANALYT SCI,GEELONG,VIC 3217,AUSTRALIA
关键词
D O I
10.1016/0375-6742(92)90025-4
中图分类号
P3 [地球物理学]; P59 [地球化学];
学科分类号
0708 ; 070902 ;
摘要
A portable, battery-operated electrochemical instrument which enables the method of staircase voltammetry to be implemented may be coupled with the method of abrasive stripping voltammetry for the direct identification of minerals in the field. Small amounts of solid mineral samples are mechanically transferred to a paraffin-impregnated graphite electrode by rubbing the electrode against the crushed mineral. The electrode is then transferred to a conventional electrochemical cell containing electrolyte, and experiments analogous to the well known method of stripping voltammetry are then completed to provide an "electrochemical spectrum" or fingerprint of the mineral. Since the technique can be applied directly with battery-operated equipment, the method can be used by geochemists and mineralogists undertaking field work in areas remote from sophisticated laboratories and mains-powered instruments normally used to characterize minerals. The method has been successfully applied to the identification of minerals such as galena, stibnite, cuprite, pyrite and chalcopyrite, although further work is required to investigate any limitations that might apply with complex mixtures or minerals, and, at the present stage of development, the method is qualitative rather than quantitative.
引用
收藏
页码:227 / 235
页数:9
相关论文
共 12 条
[1]  
[Anonymous], INVERSE POLAROGRAPHI
[2]  
[Anonymous], 1985, STRIPPING ANAL PRINC
[3]   A BATTERY-POWERED, COMPUTERIZED INSTRUMENT FOR THE FIELD-BASED DETERMINATION OF FLUORIDE BY THE ION-SELECTIVE ELECTRODE TECHNIQUE [J].
BOND, AM ;
HUDSON, HA ;
MOSBY, V ;
TAN, SN ;
WALTER, FL .
ANALYTICA CHIMICA ACTA, 1988, 208 (1-2) :195-206
[4]   BATTERY-OPERATED MICROCOMPUTER-BASED ELECTROCHEMICAL INSTRUMENTATION [J].
BOND, AM ;
HUDSON, HA ;
TAN, SN ;
WALTER, FL .
TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 1988, 7 (05) :159-164
[5]   A BATTERY-POWERED, MICROPROCESSOR-CONTROLLED, PROGRAMMABLE FUNCTION GENERATOR FOR FIELD-BASED STRIPPING VOLTAMMETRY WITH CONVENTIONAL AND MICROSIZED ELECTRODES [J].
BOND, AM ;
LUSCOMBE, DL ;
TAN, SN ;
WALTER, FL .
ELECTROANALYSIS, 1990, 2 (03) :195-202
[6]   CALCULATION OF THERMODYNAMIC DATA FROM VOLTAMMETRY OF SOLID LEAD AND MERCURY DITHIOCARBAMATE COMPLEXES MECHANICALLY ATTACHED TO A GRAPHITE ELECTRODE [J].
BOND, AM ;
SCHOLZ, F .
JOURNAL OF PHYSICAL CHEMISTRY, 1991, 95 (19) :7460-7465
[7]  
BOND AM, 1991, IN PRESS LANGMUIR
[8]   FIELD ANALYSIS OF HEAVY-METALS BY PORTABLE DIGITAL VOLTAMMETER [J].
MANN, AW ;
LINTERN, MJ .
JOURNAL OF GEOCHEMICAL EXPLORATION, 1984, 22 (1-3) :333-348
[9]   IDENTIFICATION OF SOLID MATERIALS WITH A NEW ELECTROCHEMICAL TECHNIQUE - THE ABRASIVE STRIPPING ANALYSIS [J].
SCHOLZ, F ;
NITSCHKE, L ;
HENRION, G .
FRESENIUS ZEITSCHRIFT FUR ANALYTISCHE CHEMIE, 1989, 334 (01) :56-58
[10]   A NEW PROCEDURE FOR FAST ELECTROCHEMICAL ANALYSIS OF SOLID MATERIALS [J].
SCHOLZ, F ;
NITSCHKE, L ;
HENRION, G .
NATURWISSENSCHAFTEN, 1989, 76 (02) :71-72