RETICULOCYTE COUNTING IN THALASSEMIC AND OTHER CONDITIONS WITH THE R-1000 SYSMEX ANALYZER

被引:25
作者
PATERAKIS, GS [1 ]
VOSKARIDOU, E [1 ]
LOUTRADI, A [1 ]
ROMBOS, J [1 ]
LOUKOPOULOS, D [1 ]
机构
[1] LAIKON GEN HOSP, THALASSEMIA UNIT, GR-11527 ATHENS, GREECE
关键词
RETICULOCYTES; AUTOMATIC COUNTERS; THALASSEMIA; SICKLE CELL SYNDROMES;
D O I
10.1007/BF01703447
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Precise reticulocyte counts are difficult to obtain by the manual method when their percentage in the blood is low or normal. In these instances, rapid reticulocyte counting by flow cytometry appears to offer more accuracy and precision. The purpose of this study was to establish reticulocyte counts in heterozygous beta-thalassemia for reference purposes and to evaluate the performance of the recently introduced apparatus R-1000 (Sysmex) in the very heterogeneous thalassemic and sickle-cell syndromes. We studied a total of 364 samples; 102 heterozygous beta-thalassemia carriers, 180 normal matched controls, 36 patients with thalassemia major or intermedia, and 46 patients with various sickle-cell syndromes. Reticulocyte counts (both as percentage and as total number) were higher in heterozygous beta-thalassemia than in normal controls (p < 0.001) and showed an inverse correlation with the respective hemoglobin values (p < 0.001). These results confirm the proposed slightly increased erythropoietic activity in heterozygous beta-thalassemia carriers. A drawback of the technique is that the reticulocyte-platelet discrimination error is signaled frequently in all conditions displaying a marked red cell heterogeneity, especially when these are associated with high reticulocyte numbers. This calls probably for readjustment of the corresponding algorithm. In addition, all these conditions show a significantly increased auramine-O mature red-cell nonspecific fluorescence.
引用
收藏
页码:218 / 222
页数:5
相关论文
共 20 条
  • [1] ALESSI M, 1990, ANN NY ACAD SCI, V612, P534
  • [2] ENUMERATION OF RETICULOCYTES USING FLUORESCENCE-ACTIVATED FLOW-CYTOMETRY
    CORASH, L
    RHEINSCHMIDT, M
    LIEU, S
    MEERS, P
    BREW, E
    [J]. PATHOLOGY AND IMMUNOPATHOLOGY RESEARCH, 1988, 7 (05): : 381 - 394
  • [3] DAVIS BH, 1989, ARCH PATHOL LAB MED, V113, P684
  • [4] EVALUATION OF RETICULOCYTE COUNTS BY FLOW-CYTOMETRY IN A ROUTINE LABORATORY
    FERGUSON, DJ
    LEE, SF
    GORDON, PA
    [J]. AMERICAN JOURNAL OF HEMATOLOGY, 1990, 33 (01) : 13 - 17
  • [5] AN AUTOMATED OPTOELECTRONIC RETICULOCYTE COUNTER
    KOJIMA, K
    NIRI, M
    SETOGUCHI, K
    TSUDA, I
    TATSUMI, N
    [J]. AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1989, 92 (01) : 57 - 61
  • [6] LAHARRAGUE P, 1990, ANN BIOL CLIN-PARIS, V48, P253
  • [7] THIAZOLE ORANGE - A NEW DYE FOR RETICULOCYTE ANALYSIS
    LEE, LG
    CHEN, CH
    CHIU, LA
    [J]. CYTOMETRY, 1986, 7 (06): : 508 - 517
  • [8] MAKRIS PE, 1982, NOUV REV FR HEMATOL, V24, P301
  • [9] CLINICAL AND HEMATOLOGICAL DATA IN 254 CASES OF BETA-THALASSEMIA TRAIT IN ITALY
    MAZZA, U
    SAGLIO, G
    CAPPIO, FC
    CAMASCHELLA, C
    NERETTO, G
    GALLO, E
    [J]. BRITISH JOURNAL OF HAEMATOLOGY, 1976, 33 (01) : 91 - 99
  • [10] METZGER DK, 1987, ARCH PATHOL LAB MED, V111, P540