CORRECTION FACTORS FOR CALIBRATION OF PLANE-PARALLEL IONIZATION CHAMBERS WITH A CO-60 GAMMA-RAY BEAM

被引:11
作者
LAITANO, RF [1 ]
GUERRA, AS [1 ]
PIMPINELLA, M [1 ]
NYSTROM, H [1 ]
KARLSSON, M [1 ]
SVENSSON, H [1 ]
机构
[1] UMEA UNIV, DEPT RADIAT PHYS, S-90185 UMEA, SWEDEN
关键词
D O I
10.1088/0031-9155/38/1/004
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The appropriate correction factors have been determined to enable plane-parallel ionization chambers to be calibrated using a Co-60 photon beam with a known air kerma rate. These factors refer to different calibration conditions and to different types of plane-parallel chambers. The main purpose of these correction factors is to allow calibration procedures more widely practicable than those based on the use of an electron beam of sufficiently high energy. The condition required to apply the correction factors is that the characteristics of the chambers to be calibrated are the same as those reported for the chambers considered in this investigation. To this end the types of chambers investigated were among the most widely used plane-parallel chambers commercially available. The results of this work also give indications on the properties of the plane-parallel chambers with regard to the chamber-phantom matching.
引用
收藏
页码:39 / 54
页数:16
相关论文
共 10 条
[1]  
[Anonymous], 1983, MED PHYS, V10, P741
[2]  
[Anonymous], 1987, TECHNICAL REPORT SER, V277
[4]  
Johansson K.-A., 1978, National and International Standardization of Radiation Dosimetry, P243
[5]   NGAS DETERMINATION FOR A PARALLEL-PLATE ION-CHAMBER [J].
KRITHIVAS, G ;
RAO, SN .
MEDICAL PHYSICS, 1986, 13 (05) :674-677
[6]   N-GAS VALUES OF THE MEMORIAL PARALLEL-PLATE CHAMBERS DETERMINED IN CO-60 AND HIGH-ENERGY ELECTRON-BEAMS [J].
KUBO, H .
MEDICAL PHYSICS, 1991, 18 (04) :749-752
[7]   DETERMINATIONS OF NGAS AND PREPL FACTORS FROM COMMERCIALLY AVAILABLE PARALLEL-PLATE CHAMBERS - AAPM TASK GROUP-21 PROTOCOL [J].
KUBO, H ;
KENT, LJ ;
KRITHIVAS, G .
MEDICAL PHYSICS, 1986, 13 (06) :908-912
[8]   CALIBRATION AND USE OF PLANE-PARALLEL IONIZATION CHAMBERS FOR THE DETERMINATION OF ABSORBED DOSE IN ELECTRON-BEAMS [J].
MATTSSON, LO ;
JOHANSSON, KA ;
SVENSSON, H .
ACTA RADIOLOGICA ONCOLOGY, 1981, 20 (06) :385-399
[9]   MEASUREMENT OF IONIZATION IN WATER, POLYSTYRENE AND A SOLID WATER PHANTOM MATERIAL FOR ELECTRON-BEAMS [J].
THWAITES, DI .
PHYSICS IN MEDICINE AND BIOLOGY, 1985, 30 (01) :41-53
[10]  
WITTKAMPER FW, 1990, 9TH ESTRO ANN M MONT, P214