DISPOSABLE ENDOSCOPIC BIOPSY FORCEPS - COMPARISON WITH STANDARD FORCEPS OF SAMPLE-SIZE AND ADEQUACY OF SPECIMEN

被引:19
作者
TURK, DJ [1 ]
KOZAREK, RA [1 ]
BOTOMAN, VA [1 ]
PATTERSON, DJ [1 ]
BALL, TJ [1 ]
机构
[1] VIRGINIA MASON MED CTR,THERAPEUT ENDOSCOPY SECT,POB 900,SEATTLE,WA 98111
关键词
BIOPSY; FORCEPS; HISTOLOGY; ENDOSCOPY;
D O I
10.1097/00004836-199102000-00016
中图分类号
R57 [消化系及腹部疾病];
学科分类号
摘要
Disposable biopsy forceps have recently been introduced into the field of endoscopy. We have analyzed biopsy size and histologic interpretation of samples obtained with the disposable forceps and compared them to those obtained with reusable forceps. In the 18 patients studied, 49 samples were collected with the reusable forceps and 47 samples with the disposable. Biopsy sites included the colorectum in 50%, esophagus in 22%, small bowel in 17%, and stomach in 11% of the patients. We found that statistically smaller samples were collected by the disposable biopsy forceps than by the reusable (2.48 +/- 1.11 mm versus 1.99 +/- 0.55 mm, p = 0.006). The smaller biopsy size with disposable forceps was not clinically important since all but one of 47 specimens were interpreted as adequate for histologic diagnosis. The convenience, potential cost savings, and prevention of the spread of communicable agents afforded by disposable biopsy forceps make them a possible alternative to conventional forcepts in some clinical settings.
引用
收藏
页码:76 / 78
页数:3
相关论文
共 7 条
  • [1] PSEUDOMONAS INFECTION OF THE BILIARY SYSTEM RESULTING FROM USE OF A CONTAMINATED ENDOSCOPE
    ALLEN, JI
    ALLEN, MO
    OLSON, MM
    GERDING, DN
    SHANHOLTZER, CJ
    MEIER, PB
    VENNES, JA
    SILVIS, SE
    [J]. GASTROENTEROLOGY, 1987, 92 (03) : 759 - 763
  • [2] BACTEREMIA WITH GASTROINTESTINAL ENDOSCOPIC PROCEDURES
    BOTOMAN, VA
    SURAWICZ, CM
    [J]. GASTROINTESTINAL ENDOSCOPY, 1986, 32 (05) : 342 - 346
  • [3] BACTEREMIA DURING ESOPHAGEAL VARICEAL SCLEROTHERAPY - ITS CAUSE AND PREVENTION
    BRAYKO, CM
    KOZAREK, RA
    SANOWSKI, RA
    TESTA, AW
    [J]. GASTROINTESTINAL ENDOSCOPY, 1985, 31 (01) : 10 - 12
  • [4] PROSPECTIVE ASSESSMENT OF RISK OF BACTEREMIA WITH COLONOSCOPY AND POLYPECTOMY
    LOW, DE
    SHOENUT, JP
    KENNEDY, JK
    SHARMA, GP
    HARDING, GKM
    DENBOER, B
    MICFLIKIER, AB
    [J]. DIGESTIVE DISEASES AND SCIENCES, 1987, 32 (11) : 1239 - 1243
  • [5] GASTROINTESTINAL ENDOSCOPY - INFECTION AND DISINFECTION
    OCONNOR, HJ
    AXON, ATR
    [J]. GUT, 1983, 24 (11) : 1067 - 1077
  • [6] CLEANING AND DISINFECTION OF EQUIPMENT FOR GASTROINTESTINAL FLEXIBLE ENDOSCOPY - INTERIM RECOMMENDATIONS OF A WORKING PARTY OF THE BRITISH-SOCIETY-OF-GASTROENTEROLOGY
    WELLER, IVD
    WILLIAMS, CB
    JEFFRIES, DJ
    GAZZARD, BG
    AXON, ATR
    HANSON, PJV
    AYLIFFE, G
    BARRISON, IG
    NEUMANN, C
    [J]. GUT, 1988, 29 (08) : 1134 - 1151
  • [7] YANG R, 1989, AM J GASTROENTEROL, V84, P1199