QUICK MEASUREMENT OF NMR-COIL SENSITIVITY WITH A DUAL-LOOP PROBE

被引:72
作者
DARRASSE, L
KASSAB, G
机构
[1] Institut d'Electronique Fondamentale (CNRS, URA 22), Université de Paris-Sud, Bâtiment 220
关键词
D O I
10.1063/1.1144020
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The radio-frequency coil sensitivity for NMR signal detection can be defined as B1/square-root P, which is the magnetic field that the coil induces at a given point per unit supplied power. We propose a new method, based on the reciprocity theorem, to quickly evaluate the sensitivity of a tuned coil. The only requirement of the method is a voltage gain measurement with a small dual-loop probe, without mobilization of the NMR unit. The probe can be used to gauge and map the sensitivity of the coil with good accuracy, while monitoring its tuning frequency omega0 and its quality factor Q. Our method is particularly convenient for development and maintenance of coils for MR imaging, and more generally could be applied to other fields involving radio-frequency characterization of inductive circuits.
引用
收藏
页码:1841 / 1844
页数:4
相关论文
共 7 条
[1]  
CHEN CN, 1989, MED SCI SERIES, P152
[2]   SIGNAL-TO-NOISE RATIO OF NUCLEAR MAGNETIC-RESONANCE EXPERIMENT [J].
HOULT, DI ;
RICHARDS, RE .
JOURNAL OF MAGNETIC RESONANCE, 1976, 24 (01) :71-85
[3]   THE SLOTTED CYLINDER - AN EFFICIENT PROBE FOR NMR IMAGING [J].
LEROYWILLIG, A ;
DARRASSE, L ;
TAQUIN, J ;
SAUZADE, M .
MAGNETIC RESONANCE IN MEDICINE, 1985, 2 (01) :20-28
[4]  
LORRAIN P, 1970, ELECTROMAGNETIC FIEL, P629
[5]   THE NMR PHASED-ARRAY [J].
ROEMER, PB ;
EDELSTEIN, WA ;
HAYES, CE ;
SOUZA, SP ;
MUELLER, OM .
MAGNETIC RESONANCE IN MEDICINE, 1990, 16 (02) :192-225
[6]  
TERMAN F, 1943, RADIO ENG HDB, P149
[7]  
1992, 11TH SMRM ANN M BERL, P4014