EVALUATION OF KEELER PULSAIR NONCONTACT TONOMETRY - RELIABILITY AND REPRODUCIBILITY

被引:15
作者
BONOMI, L [1 ]
BARAVELLI, S [1 ]
COBBE, C [1 ]
TOMAZZOLI, L [1 ]
机构
[1] UNIV VERONA,INST OPHTHALMOL,CTR OSPED BG TRENTO,I-37126 VERONA,ITALY
关键词
D O I
10.1007/BF00167869
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Intraocular pressure was evaluated in 414 eyes using Keeler Pulsair tonometry in comparison with values obtained by Goldmann tonometry. No statistically significant differences were found between the two techniques. Landis K statistical analysis was used to evaluate the measurements obtained by two operators in 125 eyes, with the results indicating good reproductibility. In 20 eyes, a statistically significant difference was found between values obtained with patients in the seated vs supine positions, whereas measurements made with subjects' heads in a supported or unsupported position yielded no statistically different values.
引用
收藏
页码:210 / 212
页数:3
相关论文
共 11 条
[1]  
BUSCEMI M, 1989, Annali di Ottalmologia e Clinica Oculistica, V115, P309
[2]  
COBBE C, 1990, IN PRESS B OCUL, V69
[3]  
DIBARI A, 1988, B OCUL, V67, P157
[4]  
DUCREY N, 1975, OPHTHALMOLOGICA, V173, P446
[5]   A NEW HAND-HELD AIR IMPULSE TONOMETER [J].
FISHER, JH ;
WATSON, PG ;
SPAETH, G .
EYE, 1988, 2 :238-242
[6]   NONCONTACT APPLANATION TONOMETER - DESCRIPTION AND CLINICAL EVALUATION [J].
FORBES, M ;
PICO, G ;
GROLMAN, B .
ARCHIVES OF OPHTHALMOLOGY, 1974, 91 (02) :134-140
[7]  
INSLER MS, 1987, ARCH OPHTHALMOL-CHIC, V105, P1358
[8]  
SORENSEN PN, 1975, ACTA OPHTHALMOL, V53, P513
[9]  
SPONSEL WE, 1989, ACTA OPHTHALMOL, V67, P567
[10]   NON-CONTACT TONOMETRY IN THE POSTOPERATIVE EYE [J].
VERNON, SA .
BRITISH JOURNAL OF OPHTHALMOLOGY, 1989, 73 (04) :247-249