ULTRAHIGH-RESOLUTION FOURIER-TRANSFORM ION-CYCLOTRON RESONANCE MASS-SPECTROMETER

被引:18
作者
ALBER, GM
MARSHALL, AG
HILL, NC
SCHWEIKHARD, L
RICCA, TL
机构
[1] OHIO STATE UNIV,DEPT CHEM,COLUMBUS,OH 43210
[2] TOSOH SMD INC,GROVE,OH 43123
[3] UNIV MAINZ,INST PHYS,W-6500 MAINZ,GERMANY
[4] OHIO STATE UNIV,CTR CAMPUS CHEM INSTRUMENT,COLUMBUS,OH 43210
关键词
D O I
10.1063/1.1144021
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A new Fourier transform ion cyclotron resonance mass spectrometer is described in three sections: magnet, vacuum system, and electronics/data system. Each component is described in detail (e.g., manufacturer's part numbers, drawings, circuit schematics, etc.). Of special interest is the high-frequency signal generation and processing required to extend the lower-mass limit for singly charged ions to 1 u. In particular, two new multiple heterodyne techniques for high-frequency measurements are described, as well as a method for producing a long-duration electron beam, and the implementation of z-axis ejection for selective removal of unwanted ions of various mass-to-charge ratios.
引用
收藏
页码:1845 / 1852
页数:8
相关论文
共 37 条