VARIABILITY IN REPROCESSING POLICIES AND PROCEDURES FOR FLEXIBLE FIBEROPTIC ENDOSCOPES IN MASSACHUSETTS HOSPITALS

被引:38
作者
REYNOLDS, CD [1 ]
RHINEHART, E [1 ]
DREYER, P [1 ]
GOLDMANN, DA [1 ]
机构
[1] MASSACHUSETTS DEPT PUBL HLTH,DIV HLTH CARE QUAL,BOSTON,MA 02130
关键词
D O I
10.1016/S0196-6553(05)80231-7
中图分类号
R1 [预防医学、卫生学];
学科分类号
1004 ; 120402 ;
摘要
Background: Concern is increasing regarding the risk of transmission of blood-borne pathogens by means of improperly reprocessed medical devices. Methods: On-site surveys of policies and practices were performed in 18 reprocessing areas in eight randomly selected hospitals in Massachusetts to assess current practices for reprocessing of flexible fiberoptic endoscopes. Manufacturers' and internal written protocols, were reviewed, employees were interviewed, and procedures were observed. Reprocessing included high-level disinfection in 17 of 18 areas, with 16 areas using 2% glutaraldehyde; ethylene oxide gas sterilization was used in one area. Results: Considerable interhospital and intrahospital variability was found in high-level disinfection procedures, including equipment, contact time (range 10 to 45 minutes), disinfectant testing, and rinsing. Disinfection of internal channels was inadequate in three areas, recommended sterilization of biopsy forceps was not performed in five, and written protocols were unavailable in three. Ad hoc deviation from established written or verbal protocols occurred in eight areas during reprocessing of flexible fiberoptic endoscopes from patients known to have AIDS; ethylene oxide sterilization was used in seven areas and a separate device was used in one. Interviews with personnel revealed that lack of knowledge of high-level disinfection contributed to the discrepancies between policy and practice. Conclusions: We conclude that reprocessing of flexible fiberoptic endoscopes is inconsistent and potentially ineffective. Knowledge that a flexible fiberoptic endoscope was used for a patient with AIDS influences practice.
引用
收藏
页码:283 / 290
页数:8
相关论文
共 20 条
[1]  
Dean, Transmission of Salmonella typhi by fiberoptic endoscopy [Letter], Lancet, 2, (1977)
[2]  
Dwyer, Klein, Istre, Robinson, Neumann, McCoy, Salmonella newport infection transmitted by fiberoptic colonoscopy, Gastrointest Endosc, 33, pp. 84-87, (1987)
[3]  
Classen, Jacobsen, Burke, Jacobsen, Evans, Serious Pseudomonas infections associated with endoscopic retrograde cholangiopancreatography, Am J Med, 84, pp. 590-596, (1988)
[4]  
Earnshaw, Clark, Thom, Outbreak of Pseudomonas aeruginosa following endoscopic retrograde cholangiopancreatography, J Hosp Infect, 6, pp. 95-97, (1985)
[5]  
Centers for Disease Control, Nosocomial infection and pseudoinfection from contaminated endoscopes and bronchoscopes—Wisconsin and Missouri, MMWR, 40, pp. 675-678, (1991)
[6]  
Leers, Disinfecting endoscopes: how not to transmit Mycobacteria tuberculosis by bronchoscopy, Can Med Assoc J, 123, pp. 275-280, (1980)
[7]  
Webb, Vall-Spinosa, Outbreak of Serratia marcescens associated with the flexible fiberbronchoscope, Chest, 68, pp. 75-78, (1975)
[8]  
Birnie, Quigley, Clements, Follet, Watkinson, Endoscopic transmission of hepatitis B, Gut, 24, pp. 171-174, (1983)
[9]  
Stryker, Gunn, Francis, Outbreak of hepatitis B associated with acupuncture, J Fam Pract, 22, pp. 155-158, (1986)
[10]  
Massachusetts Department of Public Health, AIDS News-letter, 5, (1989)