CROSS CALIBRATION OF QDR-2000 AND QDR-1000 DUAL-ENERGY X-RAY DENSITOMETERS FOR BONE-MINERAL AND SOFT-TISSUE MEASUREMENTS

被引:105
作者
ABRAHAMSEN, B
GRAM, J
HANSEN, TB
BECKNIELSEN, H
机构
[1] Department of Medical Endocrinology M, Odense University Hospital, Odense
关键词
BONE; DENSITOMETRY; DUAL-ENERGY X-RAY ABSORPTIOMETRY; BONE MINERAL DENSITY; LONGITUDINAL STUDIES; PRECISION;
D O I
10.1016/8756-3282(94)00054-9
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Replacement of dual energy X-ray densitometry (DXA) equipment may be necessary in many labs with time, or new equipment may be added. In this context we compared patient scans between a Hologic QDR-1000W and a QDR-2000 (n = 29-43, depending on anatomic region) and between QDR-2000 single beam (SE) and fan beam (FB) modes (n = 40-62) as a quality control measure. A total of 83 subjects (79 females and four males) with a wide range of bone mineral densities (BMD) were studied. There was a linear relationship between results with the QDR-1000W and QDR-2000 in SE mode, and between SE and FB mode on the QDR-2000, but the magnitude of the coefficients and constants differed for the different anatomic regions. In SE mode the QDR-2000 underestimated whole body and forearm BMD by 3% relative to the QDR-1000W, even after cross calibration using a spine phantom. Femoral total BMD was slightly, but not significantly, underestimated. Thus, additional postanalysis correction had to be applied to forearm and whole-body scans. In FB with the QDR-2000, spine and. whole-body BMD was underestimated by 3%, but femur total and neck BMD was overestimated by 2.2 and 2.8%, respectively, compared with SE scans on the same device. Soft-tissue composition with FB (enhanced analysis protocol) on the QDR-2000 differed greatly from that obtained using SE (standard protocol), Lean tissue mass was 4 kg lower and fat mass 4 kg higher in FB mode. Also, total body BMC was 7% lower using FB. Patients should be followed using the same scanning mode and analysis protocol whenever possible. Cross calibration by means of a single BMD anthropomorphic spine phantom is not sufficient to ensure identical results in all anatomic regions when a QDR-2000 replaces a QDR-1000W or is switched to FB mode. Linear regressions in vivo should be determined for each anatomic region.
引用
收藏
页码:385 / 390
页数:6
相关论文
共 10 条
  • [1] DUAL X-RAY ABSORPTIOMETRY - A COMPARISON BETWEEN FAN-BEAM AND PENCIL BEAM SCANS
    BLAKE, GM
    PARKER, JC
    BUXTON, FMA
    FOGELMAN, I
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1993, 66 (790) : 902 - 906
  • [2] X-RAY DUAL-PHOTON ABSORPTIOMETRY - A NEW METHOD FOR THE MEASUREMENT OF BONE-DENSITY
    CULLUM, ID
    ELL, PJ
    RYDER, JP
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1989, 62 (739) : 587 - 592
  • [3] DRINKWATER DT, 1993, J BONE MINER RES, V8, pS352
  • [4] DUNN W L, 1992, Journal of Nuclear Medicine, V33, P1063
  • [5] EIKEN P, 1994, IN PRESS BONE
  • [6] CROSS-CALIBRATION OF DXA EQUIPMENT - UPGRADING FROM A HOLOGIC QDR 1000/W TO A QDR-2000
    FAULKNER, KG
    GLUER, CC
    ESTILO, M
    GENANT, HK
    [J]. CALCIFIED TISSUE INTERNATIONAL, 1993, 52 (02) : 79 - 84
  • [7] HARPER KD, 1992, J BONE MINER RES, V7, pS19
  • [8] Lilley J, 1991, Osteoporos Int, V1, P141, DOI 10.1007/BF01625443
  • [9] ORWOLL ES, 1991, J BONE MINER RES, V6, P191
  • [10] PRITCHARD JE, 1993, EUR J CLIN NUTR, V47, P216