INTRAARTICULAR REACTION ASSOCIATED WITH THE USE OF FREEZE-DRIED, ETHYLENE OXIDE-STERILIZED BONE-PATELLA TENDON-BONE ALLOGRAFTS IN THE RECONSTRUCTION OF THE ANTERIOR CRUCIATE LIGAMENT

被引:255
作者
JACKSON, DW
WINDLER, GE
SIMON, TM
机构
[1] South. California Center, Sports Medicine, Long Beach, CA 90806
关键词
D O I
10.1177/036354659001800101
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
One hundred nine patients over a 3 year period underwent reconstruction for chronic ACL ruptures using a freeze-dried, ethylene oxide-sterilized bone-patella tendon-bone allograft. Seven patients (6.4%) developed a characteristic persistent intraarticular reaction. This reaction was characterized by persistent synovial effusion with collagenous particulates and cellular inflammatory response. Synovial biopsies in all cases showed a similar chronic inflammatory process, characterized by fibrin, collagen, and phagocytic cells. The intraarticular white cells were preominantly lymphocytes. Removal of the allograft resulted in resolution of the reaction in all of the patients. Three of the seven patients showed HLA conversion. Gas chromatography demonstrated detectable levels of ethylene chlorohydrin, a toxic reaction product of ethylene oxide, within the allograft and synovium 14 months following implantation of one graft. These seven cases presented strongly suggest a nonspecific or immune mediated response that must be further delineated. The use of freeze-dried, ethylene oxide-sterilized allografts using standard techniques cannot be recommended for reconstruction of the ACL.