IN-VIVO CORROSION OF MODULAR HIP-PROSTHESIS COMPONENTS IN MIXED AND SIMILAR METAL COMBINATIONS - THE EFFECT OF CREVICE, STRESS, MOTION, AND ALLOY COUPLING

被引:473
作者
GILBERT, JL
BUCKLEY, CA
JACOBS, JJ
机构
[1] RUSH PRESBYTERIAN ST LUKES MED CTR,DEPT ORTHOPED SURG,CHICAGO,IL 60612
[2] NORTHWESTERN UNIV,DEPT BIOMED ENGN,CHICAGO,IL 60611
来源
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH | 1993年 / 27卷 / 12期
关键词
D O I
10.1002/jbm.820271210
中图分类号
R318 [生物医学工程];
学科分类号
0831 [生物医学工程];
摘要
One hundred forty-eight retrieved modular hip prostheses of both mixed (Ti-6Al-4V/Co-Cr) and similar (Co-Cr/Co-Cr) metal combinations were examined and positive evidence of corrosive attack was found in the conical taper region between head and stem. Significant corrosion was observed in both mixed and similar metal combinations with 16% of necks and 35% of heads (for mixed-metal cases), and 14% of necks and 23% of heads (for similar-metal cases) showing moderate to severe corrosive attack. There was a significant correlation between the percentage of prostheses with moderate to severe corrosion and the duration of implantation for both mixed and similar metal cases, indicating that this corrosion process is progressive in time. Moderate to severe corrosion was seen as early as 2.5 and 11 months (mixed and similar metals, respectively). Scanning electron microscopy and x-ray analysis identified several forms of corrosive attack in the cobalt-based component of the taper. These included, for both mixed and same metal combinations: preferential dissolution of cobalt, fretting, and pitting; mixed metals only: the formation of a Ti-Cr-Mo interfacial phase and interdendritic corrosion; and for similar metals: intergranular attack adjacent to grain boundaries enriched in molybdenum and silicon. It is hypothesized that the restricted crevice environment, coupled with high cyclic stresses which cause repeated fracture of the passive oxide films in the taper, result in an unstable electrochemical environment within the crevice for both the cobalt alloy and Ti-alloy passive films. The passivity of these alloys is subsequently lost and active attack of the taper results. Also, the repeated fracturing of the passive films will result in large amounts of corrosion products being formed. This corrosion and particulate accumulation could result in loss of mechanical integrity of the implants in vivo, create particles for third body wear, and release particles into the surrounding tissues. (C) 1993 John Wiley and Sons, Inc.
引用
收藏
页码:1533 / 1544
页数:12
相关论文
共 32 条
[1]
METALLIC WEAR IN FAILED TITANIUM-ALLOY TOTAL HIP REPLACEMENTS - A HISTOLOGICAL AND QUANTITATIVE-ANALYSIS [J].
AGINS, HJ ;
ALCOCK, NW ;
BANSAL, M ;
SALVATI, EA ;
WILSON, PD ;
PELLICCI, PM ;
BULLOUGH, PG .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1988, 70A (03) :347-356
[2]
DOES CORROSION MATTER [J].
BLACK, J .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1988, 70 (04) :517-520
[3]
CREVICE AND FRETTING CORROSION OF STAINLESS-STEEL PLATES AND SCREWS [J].
BROWN, SA ;
SIMPSON, JP .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1981, 15 (06) :867-878
[4]
BROWN SA, 1992, T SOC BIOMATERIALS I, V15, P59
[5]
BUCKLEY CA, 1992, T SOC BIOMATER IMPLA, V15, P58
[6]
Cohen J, 1968, Clin Orthop Relat Res, V61, P167
[7]
CORROSION BETWEEN THE COMPONENTS OF MODULAR FEMORAL HIP PROSTHESES [J].
COLLIER, JP ;
SURPRENANT, VA ;
JENSEN, RE ;
MAYOR, MB ;
SURPRENANT, HP .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1992, 74 (04) :511-517
[8]
COLLIER JP, 1991, CLIN ORTHOP RELAT R, V271, P305
[9]
DEVINE TM, 1974, THESIS MIT
[10]
DUJOVNE AR, 1992, T WORLD BIOMATER C, V4, P264