SUBSAMPLE VOLUME PROCESSING OF DOPPLER ULTRASOUND SIGNALS

被引:21
作者
HOEKS, APG [1 ]
BRANDS, PJ [1 ]
ARTS, TGJ [1 ]
RENEMAN, RS [1 ]
机构
[1] UNIV LIMBURG,CARDIOVASC RES INST MAASTRICHT,DEPT PHYSIOL,6200 MD MAASTRICHT,NETHERLANDS
关键词
ACCURACY; PULSED DOPPLER; RF PROCESSING; SIGNAL SIMULATION; SUBSAMPLE VOLUMES; ULTRASOUND; VELOCITY ESTIMATION; VARIANCE;
D O I
10.1016/0301-5629(94)90054-X
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Processing of Doppler signals produced by pulsed Doppler systems is based on the assumption that the phase of the received high frequency ultrasound signals changes linearly with depth. However, the random spatial distribution of scatterers is not in accordance with this basic assumption. Consequently, averaging af the demodulated signal over an observation window, covering a few periods of the received signal, does not improve the estimate for the instantaneous quadrature components of the Doppler signal originating from a given depth. Hence, the accuracy of the Doppler velocity estimate is independent of the length of the observation window employed. However, splitting the observation window in subsample volumes, each with a length of one period at the emission frequency, and combining the Doppler signals of the subsample volumes at the last stage of signal processing, i.e., mean Doppler frequency estimation using the autocorrelation technique, results in a consideration reduction of the variance of the velocity estimate. Using a computer simulation of the signal processing involved, it is demonstrated that with subsample volume processing the variance of the velocity estimate attains the same variance as is expected for the RF cross correlation technique.
引用
收藏
页码:953 / 965
页数:13
相关论文
共 13 条
[1]   DETERMINATION OF TISSUE MOTION VELOCITY BY CORRELATION INTERPOLATION OF PULSED ULTRASONIC ECHO SIGNALS [J].
DEJONG, PGM ;
ARTS, T ;
HOEKS, APG ;
RENEMAN, RS .
ULTRASONIC IMAGING, 1990, 12 (02) :84-98
[2]  
DEJONG PGM, 1991, THESIS U LIMBURG MAA
[3]   VOLUMETRIC BLOOD-FLOW VIA TIME-DOMAIN CORRELATION - EXPERIMENTAL-VERIFICATION [J].
EMBREE, PM ;
OBRIEN, WD .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1990, 37 (03) :176-189
[4]   THE EFFECT OF FREQUENCY-DEPENDENT SCATTERING AND ATTENUATION ON THE ESTIMATION OF BLOOD VELOCITY USING ULTRASOUND [J].
FERRARA, KW ;
ALGAZI, VR ;
LIU, JX .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1992, 39 (06) :754-767
[5]   FLOW VELOCITY PROFILE VIA TIME-DOMAIN CORRELATION ERROR ANALYSIS AND COMPUTER-SIMULATION [J].
FOSTER, SG ;
EMBREE, PM ;
OBRIEN, WD .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1990, 37 (03) :164-175
[6]   CURRENT TIME-DOMAIN METHODS FOR ASSESSING TISSUE MOTION BY ANALYSIS FROM REFLECTED ULTRASOUND ECHOES - A REVIEW [J].
HEIN, IA ;
OBRIEN, WD .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1993, 40 (02) :84-102
[7]   COMPARISON OF THE PERFORMANCE OF THE RF CROSS-CORRELATION AND DOPPLER AUTOCORRELATION TECHNIQUE TO ESTIMATE THE MEAN VELOCITY OF SIMULATED ULTRASOUND SIGNALS [J].
HOEKS, APG ;
ARTS, TGJ ;
BRANDS, PJ ;
RENEMAN, RS .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1993, 19 (09) :727-740
[8]   EFFECT OF SAMPLE WINDOW LENGTH ON THE CORRELATION BETWEEN RF SIGNAL AND PULSED DOPPLER SIGNAL INTENSITY [J].
HOEKS, APG ;
BRANDS, PJ ;
RENEMAN, RS .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1994, 20 (01) :35-40
[9]   REAL-TIME TWO-DIMENSIONAL BLOOD-FLOW IMAGING USING AN AUTO-CORRELATION TECHNIQUE [J].
KASAI, C ;
NAMEKAWA, K ;
KOYANO, A ;
OMOTO, R .
IEEE TRANSACTIONS ON SONICS AND ULTRASONICS, 1985, 32 (03) :458-464
[10]  
Peronneau P., 1974, CARDIOVASCULAR APPL, P66