摘要
This paper describes a sensor array of several amperometric electrodes, each coated with a different permselective film. Coatings with different transport properties, based on size (cellulose acetate), charge (Nation, poly(vinylpyrldine), poly(ester sulfonic acid)), and polarity (phospholipid) are employed In connection with a four-electrode thin-layer flow detector. With equipotential operation, the array's response pattern of each analyte provides a unique characterization of the Individual components. Multicomponent analysis Is obtained by taking advantage of the partial selectivity of the individual sensors and using a pattern recognition (multiple linear regression) method. Additional information is obtained by recording the complete hydrodynamic voltammogram at the Individual sensors. The merits of the new array are Illustrated for the quantification of neurologically significant catechol compounds. © 1990, American Chemical Society. All rights reserved.
引用
收藏
页码:1924 / 1927
页数:4
相关论文
共 23 条
[1]  
AMINE A, 1989, ANAL LETT, V22, P2403
[2]   SPARINGLY SELECTIVE ION-SELECTIVE ELECTRODE ARRAYS FOR MULTICOMPONENT ANALYSIS [J].
BEEBE, K ;
UERZ, D ;
SANDIFER, J ;
KOWALSKI, B .
ANALYTICAL CHEMISTRY, 1988, 60 (01) :66-71
[3]   PULSED AMPEROMETRIC DETECTION OF GLUCOSE IN BIOLOGICAL-FLUIDS AT A SURFACE-MODIFIED GOLD ELECTRODE [J].
BINDRA, DS ;
WILSON, GS .
ANALYTICAL CHEMISTRY, 1989, 61 (22) :2566-2570
[4]   SELECTION OF ADSORBATES FOR CHEMICAL SENSOR ARRAYS BY PATTERN-RECOGNITION [J].
CAREY, WP ;
BEEBE, KR ;
KOWALSKI, BR ;
ILLMAN, DL ;
HIRSCHFELD, T .
ANALYTICAL CHEMISTRY, 1986, 58 (01) :149-153
[5]   MULTICOMPONENT ANALYSIS USING AN ARRAY OF PIEZOELECTRIC CRYSTAL SENSORS [J].
CAREY, WP ;
BEEBE, KR ;
KOWALSKI, BR .
ANALYTICAL CHEMISTRY, 1987, 59 (11) :1529-1534
[6]   MONITORING A DRYER OPERATION USING AN ARRAY OF PIEZOELECTRIC-CRYSTALS [J].
CAREY, WP ;
KOWALSKI, BR .
ANALYTICAL CHEMISTRY, 1988, 60 (06) :541-544
[7]   HYDROPHOBIC STRIPPING VOLTAMMETRY USING A LIPID-MODIFIED GLASSY-CARBON ELECTRODE [J].
CHASTEL, O ;
KAUFFMANN, JM ;
PATRIARCHE, GJ ;
CHRISTIAN, GD .
ANALYTICAL CHEMISTRY, 1989, 61 (02) :170-173
[8]   ELECTRODES MODIFIED WITH A FILM OF PHOSPHATIDYLCHOLINE - ELECTROCHEMISTRY INSIDE A LIPID LAYER [J].
GARCIA, OJ ;
QUINTELA, PA ;
KAIFER, AE .
ANALYTICAL CHEMISTRY, 1989, 61 (09) :979-981
[9]   NAFION-COATED ELECTRODES WITH HIGH SELECTIVITY FOR CNS ELECTROCHEMISTRY [J].
GERHARDT, GA ;
OKE, AF ;
NAGY, G ;
MOGHADDAM, B ;
ADAMS, RN .
BRAIN RESEARCH, 1984, 290 (02) :390-395
[10]   A GLUCOSE ELECTRODE BASED ON CARBON PASTE CHEMICALLY MODIFIED WITH A FERROCENE-CONTAINING SILOXANE POLYMER AND GLUCOSE-OXIDASE, COATED WITH A POLY(ESTER-SULFONIC ACID) CATION-EXCHANGER [J].
GORTON, L ;
KARAN, HI ;
HALE, PD ;
INAGAKI, T ;
OKAMOTO, Y ;
SKOTHEIM, TA .
ANALYTICA CHIMICA ACTA, 1990, 228 (01) :23-30