AN ADAPTIVE CURRENT TOMOGRAPH USING VOLTAGE SOURCES

被引:17
作者
ZHU, QS
LIONHEART, WRB
LIDGEY, FJ
MCLEOD, CN
PAULSON, KS
PIDCOCK, MK
机构
[1] Electrical Impedance Tomography Research Group, Oxford Brookes University
关键词
D O I
10.1109/10.212056
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This paper describes the development of an adaptive electric current tomography system which contains a novel front-end analog architecture. Programmable voltage sources have been used to deliver currents into the study object and to avoid the difficulties of obtaining high quality current sources. Through inverting an admittance matrix, the system is capable of achieving a desired current drive pattern by applying a computed voltage pattern. The tomograph, operating at 9.6 kHz, comprises 32 driving electrodes and 32 voltage measurement electrodes. The study of system noise performance shows high SNR in the data acquisition which is enhanced by a digital demodulation scheme. In vitro reconstruction images have been obtained with the data collected by the tomograph.
引用
收藏
页码:163 / 168
页数:6
相关论文
共 15 条
[1]   APPLIED POTENTIAL TOMOGRAPHY [J].
BARBER, DC ;
BROWN, BH .
JOURNAL OF PHYSICS E-SCIENTIFIC INSTRUMENTS, 1984, 17 (09) :723-733
[2]   SPECTRA OF QUANTIZED SIGNALS [J].
BENNETT, WR .
BELL SYSTEM TECHNICAL JOURNAL, 1948, 27 (03) :446-472
[3]  
BRECKON WR, 1988, INFORMATION PROCESSI
[4]  
Brown B H, 1987, Clin Phys Physiol Meas, V8 Suppl A, P91, DOI 10.1088/0143-0815/8/4A/012
[5]  
BROWN BH, 1987, CLIN PHYS PHYSL M SA, V8, P119
[6]  
COOK RD, 1991, P IEEE EMBS, V13, P22
[7]  
GISSER DG, 1990, P 12 ANN INT C IEEE, V12, P112
[8]  
HOROWITZ P, 1989, ART ELECTRONICS, P428
[9]   DISTINGUISHABILITY OF CONDUCTIVITIES BY ELECTRIC-CURRENT COMPUTED-TOMOGRAPHY [J].
ISAACSON, D .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1986, 5 (02) :91-95
[10]  
Kim Y, 1987, Clin Phys Physiol Meas, V8 Suppl A, P63, DOI 10.1088/0143-0815/8/4A/008