RADIATIVE LOSSES OF A BIRDCAGE RESONATOR

被引:25
作者
HARPEN, MD
机构
关键词
RADIATIVE LOSSES; BIRDCAGE RESONATORS;
D O I
10.1002/mrm.1910290522
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
We present a derivation of the losses in a birdcage resonator due to radiation. We also present an expression for the radiation limited 0. It is shown that in head coil imaging at 63 MHz radiative losses may account for 20% of the total loss with a radiation limited 0 on the order of 150. The results are shown to be consistent with those reported in the recent literature.
引用
收藏
页码:713 / 716
页数:4
相关论文
共 8 条
[1]  
Arfken G., 1985, MATH METHODS PHYS, P573
[2]   THEORETICAL DESCRIPTION OF THE INTERACTION BETWEEN BIRDCAGE COILS AND LOSSY SAMPLES [J].
HARPEN, MD .
MEDICAL PHYSICS, 1991, 18 (05) :1052-1056
[3]   HARMONIC STRUCTURE OF CYLINDRICAL IMAGING COILS [J].
HARPEN, MD .
MEDICAL PHYSICS, 1990, 17 (04) :686-690
[4]   AN EFFICIENT, HIGHLY HOMOGENEOUS RADIOFREQUENCY COIL FOR WHOLE-BODY NMR IMAGING AT 1.5-T [J].
HAYES, CE ;
EDELSTEIN, WA ;
SCHENCK, JF ;
MUELLER, OM ;
EASH, M .
JOURNAL OF MAGNETIC RESONANCE, 1985, 63 (03) :622-628
[5]   SENSITIVITY OF THE ZEUGMATOGRAPHIC EXPERIMENT INVOLVING HUMAN SAMPLES [J].
HOULT, DI ;
LAUTERBUR, PC .
JOURNAL OF MAGNETIC RESONANCE, 1979, 34 (02) :425-433
[6]  
JACKSON JD, 1975, CLASSICAL ELECTRODYN, P7425
[7]  
JJACKSON JD, 1985, CLASSICAL ELECTRODYN, P391
[8]   A QUADRATURE COIL FOR THE ADULT HUMAN HEAD [J].
SANK, VJ ;
CHEN, CN ;
HOULT, DI .
JOURNAL OF MAGNETIC RESONANCE, 1986, 69 (02) :236-242