SPECTRAL CHARACTERIZATION OF THE EFFECT OF VISCOSITY ON FURA-2 FLUORESCENCE - EXCITATION WAVELENGTH OPTIMIZATION ABOLISHES THE VISCOSITY ARTIFACT

被引:20
作者
BUSA, WB
机构
[1] Department of Biology, The Johns Hopkins University, Baltimore, MD
关键词
D O I
10.1016/0143-4160(92)90066-2
中图分类号
Q2 [细胞生物学];
学科分类号
071009 ; 090102 ;
摘要
A systematic study of the spectral characteristics of the viscosity artifact in Fura-2 based [Ca2+] measurements reveals' that, by selecting excitation wavelengths approximately 10 nm longer than those routinely employed and modestly reducing excitation bandpasses, the magnitude of the artifact can be reduced to experimentally undetectable levels without greatly impairing [Ca2+] measurements. The feasibility of this approach was confirmed on a ratio imaging microscope; the magnitude of the artifact observed in dextran-conjugated Fura-2 solutions prepared in water or in 50% sucrose was not statistically significant using an excitation wavelength pair of 361/389 nm, whereas at 350/380 nm [Ca2+] was underestimated by 34% in the higher viscosity solution. Thus, provided potential pitfalls are taken into account, a simple change in imaging protocol can avoid the viscosity artifact without recourse to correction factors. This approach may be employed either routinely, or else merely to test whether apparent [Ca2+]i differences observed at more conventional wavelengths arise from the viscosity artifact.
引用
收藏
页码:313 / 319
页数:7
相关论文
共 5 条
[1]  
GRYNKIEWICZ G, 1985, J BIOL CHEM, V260, P3440
[2]   ALTERATION OF INTRACELLULAR FURA-2 FLUORESCENCE BY VISCOSITY - A SIMPLE CORRECTION [J].
POENIE, M .
CELL CALCIUM, 1990, 11 (2-3) :85-&
[3]   ASSESSMENT OF FURA-2 FOR MEASUREMENTS OF CYTOSOLIC FREE CALCIUM [J].
ROE, MW ;
LEMASTERS, JJ ;
HERMAN, B .
CELL CALCIUM, 1990, 11 (2-3) :63-&
[4]  
TSIEN R, 1989, METHOD ENZYMOL, V172, P230
[5]  
CRC HDB CHEM PHYSICS