REPRESENTATIVE DILUTION OF AEROSOLS BY A FACTOR OF 10,000

被引:11
作者
HELSPER, C [1 ]
MOLTER, W [1 ]
HALLER, P [1 ]
机构
[1] AC LAB SPIEZ,RUSTUNGSDIENSTE GRP,CH-3700 SPIEZ,SWITZERLAND
关键词
AEROSOL DILUTION; OPTICAL PARTICLE COUNTERS; CONDENSATION NUCLEI COUNTERS; HEPA-ULPA-FILTERS;
D O I
10.1016/0021-8502(90)90323-P
中图分类号
TQ [化学工业];
学科分类号
0817 ;
摘要
The performance of an ejector type dilution system has been investigated for mono- and polydisperse DES-particles in the size range between 0.1-mu-m and 2-mu-m using a Condensation Nuclei Counter and a Laser Particle Counter. By cascading up to four stages with a dilution factor of 10, overall dilution factors of up to 10,000 could be achieved. Within the accuracy of the measurement techniques used and within the statistical limits given by the rather low particle concentration downstream of the fourth dilution stage the dilution systems proved to work very well.
引用
收藏
页码:S637 / S640
页数:4