COMPARISON OF CYCLOSPORINE BLOOD-LEVELS MEASURED BY RADIOIMMUNOASSAY AND TDX ASSAY USING MONOCLONAL-ANTIBODIES

被引:3
作者
KLOTZ, U
机构
[1] Dr. Margarete Fischer-Bosch Institute of Clinical Pharmacology, Stuttgart
关键词
CYCLOSPORINE; BLOOD LEVELS; RADIOIMMUNOASSAY; TDX; MONOCLONAL ANTIBODIES;
D O I
10.1097/00007691-199109000-00012
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Whole blood steady-state trough concentrations of cyclosporine were measured by radioimmunoassay (RIA) and TDx assays employing monoclonal antibodies in 82 samples from 39 renal transplant patients. Intra- and interassay coefficients of variation for the new TDx technique averaged 3.2 and 5.4%, respectively. In the concentration range of 40 to 717 ng/ml, there was a linear correlation (r = 0.884, p < 0.001) between both assays [slope of 0.863; y intercept (RIA values) = 15 ng/ml], indicating that the methods can be used interchangeably. Since the automated TDx assay is faster and causes less waste problems, this reliable technique offers some definite advantages for routine clinical use.
引用
收藏
页码:461 / 463
页数:3
相关论文
共 9 条
[1]   EVALUATION OF FLUORESCENCE POLARIZATION IMMUNOASSAY FOR DETERMINATION OF CYCLOSPORINE IN PLASMA [J].
HAYASHI, Y ;
SHIBATA, N ;
MINOUCHI, T ;
SHIBATA, H ;
ONO, T ;
SHIMAKAWA, H .
THERAPEUTIC DRUG MONITORING, 1989, 11 (02) :205-209
[2]   COMPARISON OF 4 DIFFERENT CYCLOSPORINE IMMUNOASSAYS IN HEART AND KIDNEY-TRANSPLANT RECIPIENTS [J].
HESSE, CJ ;
BAAN, CC ;
DAANE, R ;
STOLK, BJM ;
JUTTE, NHPM ;
HENDRIKS, GFJ ;
WEIMAR, W .
THERAPEUTIC DRUG MONITORING, 1990, 12 (06) :547-553
[3]  
HOLT DW, 1987, SANDIMMUM MONITORING
[4]  
KAHAN BD, 1989, NEW ENGL J MED, V321, P1725
[5]  
QUESNIAUX VFJ, 1989, PHARMACOL REV, V41, P249
[6]  
SANGHVI A, 1988, CLIN CHEM, V34, P1904
[7]  
SCHROEDER TJ, 1989, THER DRUG MONIT, V11, P480
[8]  
VOGT W, 1988, CLIN CHEM, V34, P1459
[9]  
YATSCOFF RW, 1990, CLIN CHEM, V36, P1969