INVITRO ANTIBIOTIC REMOVAL AND BACTERIAL RECOVERY FROM BLOOD WITH AN ANTIBIOTIC REMOVAL DEVICE

被引:36
作者
LINDSEY, NJ [1 ]
RIELY, PE [1 ]
机构
[1] MARION LABS INC, KANSAS CITY, MO 64137 USA
关键词
D O I
10.1128/JCM.13.3.503-507.1981
中图分类号
Q93 [微生物学];
学科分类号
071005 ; 100705 ;
摘要
The antibiotic removal device manufactured by Marion Laboratories is intended for treatment, before culture, of blood specimens from hospital patients being treated with antibiotics. Measurement of 13 antibiotics showed that the antibiotic removal device removed amikacin, ampicillin, carbenicillin, cefazolin, cephalothin, chloramphenicol, gentamicin, nafcillin, tetracycline, tobramycin and vancomycin and reduced cefoxitin and ticarcillin to extremely low levels. Three combinations of antibiotics were similarly removed or reduced. Five species of anaerobic bacteria, 1 yeast species and 6 spp. of facultative or aerobic bacteria were used to challenge the possibility that the antibiotic removal device would trap or inhibit microorganisms. All were recovered from the device in the same numbers as were inoculated.
引用
收藏
页码:503 / 507
页数:5
相关论文