THE INITIATION OF FAILURE IN CEMENTED FEMORAL COMPONENTS OF HIP ARTHROPLASTIES

被引:472
作者
JASTY, M
MALONEY, WJ
BRAGDON, CR
OCONNOR, DO
HAIRE, T
HARRIS, WH
机构
[1] MASSACHUSETTS GEN HOSP, DEPT ORTHOPAED SURG, HIP & IMPLANT UNIT, BOSTON, MA 02114 USA
[2] MASSACHUSETTS GEN HOSP, DEPT ORTHOPAED SURG, ORTHOPAED BIOMECH LABS, BOSTON, MA 02114 USA
来源
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME | 1991年 / 73卷 / 04期
关键词
D O I
10.1302/0301-620X.73B4.2071634
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
We studied 16 femora retrieved at post-mortem from symptomless patients who had a satisfactory cemented total hip arthroplasty from two weeks to 17 years earlier, with the aim of delineating the initial mechanisms involved in loosening. Only one specimen showed radiographic evidence of loosening; the other 15 were stable to mechanical testing at 17.0 Nm of torque. In all 16 specimens, the cement-bone interface was intact with little fibrous tissue formation. By contrast, separation at the cement-prosthesis interface and fractures in the cement mantle were frequent. The most common early feature was debonding of the cement from the metal, seen at the proximal and distal ends of the prosthesis. Specimens which had been in place for longer also showed circumferential fractures in the cement, near the cement-metal interface, and radial fractures extending from this interface into the cement and sometimes to the bony interface. The most extensive cement fractures appeared to have started at or near sharp corners in the metal, or where the cement mantle was thin or incomplete. Fractures were also related to voids in the cement. The time relationship in this series suggested that long-term failure of the fixation of cemented femoral components was primarily mechanical, starting with debonding at the interface between the cement and the prosthesis, and continuing as slowly developing fractures in the cement mantle.
引用
收藏
页码:551 / 558
页数:8
相关论文
共 23 条
[1]  
AHMED A M, 1984, Journal of Orthopaedic Research, V2, P105, DOI 10.1002/jor.1100020201
[2]   CENTRIFUGATION AS A METHOD OF IMPROVING TENSILE AND FATIGUE PROPERTIES OF ACRYLIC BONE-CEMENT [J].
BURKE, DW ;
GATES, EI ;
HARRIS, WH .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1984, 66A (08) :1265-1273
[3]  
Charnley J., 1979, LOW FRICTION ARTHROP
[4]   ANALYSIS OF FEMORAL COMPONENT STEM DESIGN IN TOTAL HIP-ARTHROPLASTY [J].
CROWNINSHIELD, RD ;
BRAND, RA ;
JOHNSTON, RC ;
MILROY, JC .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1980, 62 (01) :68-78
[5]   CEMENT STRAIN-MEASUREMENT SURROUNDING LOOSE AND WELL-FIXED FEMORAL COMPONENT STEMS [J].
CROWNINSHIELD, RD ;
TOLBERT, JR .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1983, 17 (05) :819-828
[6]   TELEMETRIC FORCE MEASUREMENTS ACROSS THE HIP AFTER TOTAL ARTHROPLASTY [J].
DAVY, DT ;
KOTZAR, GM ;
BROWN, RH ;
HEIPLE, KG ;
GOLDBERG, VM ;
HEIPLE, KG ;
BERILLA, J ;
BURSTEIN, AH .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1988, 70A (01) :45-50
[7]   TOTAL HIP ARTHROPLASTIES IN PATIENTS LESS THAN 45-YEARS-OLD [J].
DORR, LD ;
TAKEI, GK ;
CONATY, JP .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1983, 65 (04) :474-479
[8]  
FORNASIER VL, 1976, CLIN ORTHOP RELAT R, V116, P248
[9]   FORMATION OF A SYNOVIAL-LIKE MEMBRANE AT THE BONE-CEMENT INTERFACE - ITS ROLE IN BONE-RESORPTION AND IMPLANT LOOSENING AFTER TOTAL HIP-REPLACEMENT [J].
GOLDRING, SR ;
JASTY, M ;
ROELKE, MS ;
ROURKE, CM ;
BRINGHURST, FR ;
HARRIS, WH .
ARTHRITIS AND RHEUMATISM, 1986, 29 (07) :836-842
[10]  
Gruen TA, 1979, CLIN ORTHOP RELAT R, V141, P17