NITINOL INTRAVASCULAR STENT - RESULTS OF PRECLINICAL EVALUATION

被引:113
作者
CRAGG, AH
DEJONG, SC
BARNHART, WH
LANDAS, SK
SMITH, TP
机构
[1] UNIV IOWA HOSP & CLIN,DEPT RADIOL,IOWA CITY,IA
[2] UNIV IOWA HOSP & CLIN,DEPT PATHOL,IOWA CITY,IA
关键词
ARTERIES; EXTREMITIES; GRAFTS AND PROSTHESES; SURGERY; INTERVENTIONAL PROCEDURES; EXPERIMENTAL; STENTS AND PROSTHESES;
D O I
10.1148/radiology.189.3.8234703
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE: To test an expandable nitinol intraluminal stent for hoop strength, biocompatibility, corrosion resistance, and patency. MATERIALS AND METHODS: Forty-four stents were implanted in the iliac arteries of 22 sheep. Followup was performed with angiography and histologic examination for up to 6 months. RESULTS: All but one stent remained widely patent during the follow-up period. Two stents in two sheep were placed incorrectly early in the study; these sheep were not followed up. Hoop strength was found to be superior to that of a similar, commercially available, self-expanding stent. Minimal corrosion was seen at 6 months, and the stent appears to be biocompatible. The stent has good radiopacity and deploys with minimal foreshortening. CONCLUSION The stent can be reliably and safely deployed in the vascular system. Clinical trials in humans are warranted.
引用
收藏
页码:775 / 778
页数:4
相关论文
共 14 条
  • [1] BIOCOMPATIBILITY OF NITINOL ALLOY AS AN IMPLANT MATERIAL
    CASTLEMAN, LS
    MOTZKIN, SM
    ALICANDRI, FP
    BONAWIT, VL
    JOHNSON, AA
    [J]. JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1976, 10 (05): : 695 - 731
  • [2] CASTLEMAN LS, 1981, BIOCOMPATIBILITY CLI
  • [3] NON-SURGICAL PLACEMENT OF ARTERIAL ENDOPROSTHESES - A NEW TECHNIQUE USING NITINOL WIRE
    CRAGG, A
    LUND, G
    RYSAVY, J
    CASTANEDA, F
    CASTANEDAZUNIGA, W
    AMPLATZ, K
    [J]. RADIOLOGY, 1983, 147 (01) : 261 - 263
  • [4] CRAGG AH, 1985, RADIOLOGY, V150, P45
  • [5] TRANS-LUMINAL EXPANDABLE NITINOL COIL STENT GRAFTING - PRELIMINARY-REPORT
    DOTTER, CT
    BUSCHMANN, RW
    MCKINNEY, MK
    ROSCH, J
    [J]. RADIOLOGY, 1983, 147 (01) : 259 - 260
  • [6] TRANSLUMINALLY-PLACED COILSPRING ENDARTERIAL TUBE GRAFTS - LONG-TERM PATENCY IN CANINE POPLITEAL ARTERY
    DOTTER, CT
    [J]. INVESTIGATIVE RADIOLOGY, 1969, 4 (05) : 329 - +
  • [7] EDIE JW, 1981, ANGLE ORTHOD, V51, P319
  • [8] DEVELOPMENT OF BLOOD-COAGULATION - A FETAL LAMB MODEL
    KISKER, CT
    ROBILLARD, JE
    CLARKE, WR
    [J]. PEDIATRIC RESEARCH, 1981, 15 (07) : 1045 - 1050
  • [9] LANDAS S, 1986, ULTRASTRUCT PATHOL, V9, P129
  • [10] THERMOMECHANICAL STUDY OF NI-TI ALLOYS
    LEE, JH
    PARK, JB
    ANDREASEN, GF
    LAKES, RS
    [J]. JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1988, 22 (06): : 573 - 588