DEVELOPMENT OF A LABORATORY ROBOTIC SYSTEM FOR AUTOMATED BIOANALYTICAL METHODS .1. THE DETERMINATION OF THEOPHYLLINE IN HUMAN PLASMA - A COMPARISON BETWEEN THE ROBOTIZED AND MANUAL METHOD

被引:15
作者
HEMPENIUS, J
WIELING, J
JONKMAN, JHG
DENOORD, OE
COENEGRACHT, PMJ
DOORNBOS, DA
机构
[1] PHARMA BIORES INT BV,POB 147,9400 AC ASSEN,NETHERLANDS
[2] UNIV GRONINGEN,DEPT ANALYT CHEM & TOXICOL,CHEMOMETR RES GRP,9713 AW GRONINGEN,NETHERLANDS
关键词
comparison with manual sample preparation; robot control program; Robotic sample preparation; theophylline in plasma; validation of robotic method;
D O I
10.1016/0731-7085(90)80044-P
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The quality of bioanalytical methods is often determined by the quality of sample preparation. Using a robot for sample treatment may give better results than manual sample preparation, since the robot lacks human behaviour and incidental errors that are part of it. The use of a laboratory robot has the additional advantage of giving each sample the same analytical history, resulting in better reproducibility. An automated method has been developed for the analysis of drugs in plasma using a laboratory robot. Theophylline was used as a probe drug. Sample preparation was automated with a Zymate II robot, followed by separation and quantitation on an HPLC-system. The robotic method showed a good correlation with the manual method, while sample throughput was doubled. © 1990.
引用
收藏
页码:313 / 320
页数:8
相关论文
共 18 条
[1]  
BROSEMER J, 1986, INT LAB NOV, P82
[2]   ROBOTICS FOR THE BIOANALYTICAL LABORATORY - A FLEXIBLE SYSTEM FOR THE ANALYSIS OF DRUGS IN BIOLOGICAL-FLUIDS [J].
FOUDA, HG ;
SCHNEIDER, RP .
TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 1987, 6 (06) :139-147
[3]  
HAWK GL, 1982, INT LAB SEP, P48
[4]   ONE-DIMENSIONAL KALMAN FILTER IN A ROBOT SYSTEM FOR TOTAL ANALYSIS OF SOLID DOSAGE FORMULATIONS [J].
HAYASHI, Y ;
MATSUDA, R ;
YOSHIOKA, S ;
TAKEDA, Y .
ANALYTICA CHIMICA ACTA, 1988, 209 (1-2) :45-55
[5]  
HURST WJ, 1987, LC GC, V5, P49
[6]   ROBOTICS IN THE LABORATORY [J].
ISENHOUR, TL .
JOURNAL OF CHEMICAL INFORMATION AND COMPUTER SCIENCES, 1985, 25 (03) :292-295
[7]   INFLUENCE OF SAMPLE PREPARATION AND STORAGE ON THEOPHYLLINE CONCENTRATIONS IN BIOLOGICAL-FLUIDS [J].
JONKMAN, JHG ;
FRANKE, JP ;
SCHOENMAKER, R ;
DEZEEUW, RA .
INTERNATIONAL JOURNAL OF PHARMACEUTICS, 1982, 10 (02) :177-180
[8]   DISPOSITION AND CLINICAL PHARMACOKINETICS OF MICROCRYSTALLINE THEOPHYLLINE [J].
JONKMAN, JHG ;
BERG, WC ;
SCHOENMAKER, R ;
DEZEEUW, RA ;
GREVING, JE ;
ORIE, NGM .
EUROPEAN JOURNAL OF CLINICAL PHARMACOLOGY, 1980, 17 (05) :379-384
[9]  
JONKMAN JHG, 1980, PHARM WEEKBLAD SC ED, V2, P49
[10]   ROBOTS IN FLEXIBLE ANALYSIS SYSTEMS [J].
KOOL, P ;
MICHOTTE, Y .
TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 1985, 4 (02) :44-50