TIME-RESOLVED FLOW-CYTOMETRY FOR THE MEASUREMENT OF LANTHANIDE CHELATE FLUORESCENCE .1. CONCEPT AND THEORETICAL EVALUATION

被引:19
作者
CONDRAU, MA
SCHWENDENER, RA
NIEDERER, P
ANLIKER, M
机构
[1] UNIV ZURICH, INST BIOMED ENGN, CH-8092 ZURICH, SWITZERLAND
[2] ETH ZURICH, CH-8092 ZURICH, SWITZERLAND
[3] UNIV ZURICH HOSP, DEPT INTERNAL MED, DIV ONCOL, CH-8091 ZURICH, SWITZERLAND
来源
CYTOMETRY | 1994年 / 16卷 / 03期
关键词
PHOTON COUNTING; OPTICAL SWITCH; BACKGROUND ELIMINATION; AUTOFLUORESCENCE; DETECTION LIMIT; EUROPIUM CHELATES;
D O I
10.1002/cyto.990160302
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The concept of a flow cytometer suited for the time-resolved measurement of lanthanide chelate luminescence with a decay time on the order of 10 mu s to 2 ms is presented and evaluated. The instrument proposed encompasses a continuous-wave laser for fluorescence excitation and an optical switch for the elimination of cellular autofluorescence decaying within 1 ns to 1 mu s during the luminescence detection period. The slowly decaying fluorescence is to be quantified by a photon-counting system, whereas light scatter and prompt fluorescence parameters are acquired by a conventional detection system. The detection limit of the method, in terms of the smallest detectable number of fluorescing chelates per cell, is examined. It was found to be nearly 30,000 complexes of a europium chelate with a decay time of 1.6 ms and a quantum efficiency of 17%, independent of fast decaying cellular autofluorescence or prompt dye emission intensity. The probability of cells passing through the instrument without being detected while the laser beam is turned off was estimated, and the implications for cell throughput and sorting performance of the instrument were assessed. At typical fluorescence detection intervals of 500 mu s to 1 ms, cell flow rates of 100-200 particles per second lead to detection probabilities of more than 90% and sorting purities comparable to those found in conventional fluorescence-activated cell sorting. (C) 1994 Wiley-Liss, Inc.
引用
收藏
页码:187 / 194
页数:8
相关论文
共 47 条
  • [1] A SINGLE LASER METHOD FOR SUBTRACTION OF CELL AUTOFLUORESCENCE IN FLOW-CYTOMETRY
    ALBERTI, S
    PARKS, DR
    HERZENBERG, LA
    [J]. CYTOMETRY, 1987, 8 (02): : 114 - 119
  • [2] AUTOFLUORESCENCE OF VIABLE CULTURED MAMMALIAN-CELLS
    AUBIN, JE
    [J]. JOURNAL OF HISTOCHEMISTRY & CYTOCHEMISTRY, 1979, 27 (01) : 36 - 43
  • [3] AN IMPROVED IMMUNOCYTOCHEMICAL PROCEDURE FOR HIGH-SENSITIVITY DETECTION OF INCORPORATED BROMODEOXYURIDINE
    BEISKER, W
    DOLBEARE, F
    GRAY, JW
    [J]. CYTOMETRY, 1987, 8 (02): : 235 - 239
  • [4] CELLULAR AUTOFLUORESCENCE - IS IT DUE TO FLAVINS
    BENSON, RC
    MEYER, RA
    ZARUBA, ME
    MCKHANN, GM
    [J]. JOURNAL OF HISTOCHEMISTRY & CYTOCHEMISTRY, 1979, 27 (01) : 44 - 48
  • [5] PREPARATION AND MICROSCOPIC VISUALIZATION OF MULTICOLOR LUMINESCENT IMMUNOPHOSPHORS
    BEVERLOO, HB
    VANSCHADEWIJK, A
    BONNET, J
    VANDERGEEST, R
    RUNIA, R
    VERWOERD, NP
    VROLIJK, J
    PLOEM, JS
    TANKE, HJ
    [J]. CYTOMETRY, 1992, 13 (06): : 561 - 570
  • [6] INORGANIC PHOSPHORS AS NEW LUMINESCENT LABELS FOR IMMUNOCYTOCHEMISTRY AND TIME-RESOLVED MICROSCOPY
    BEVERLOO, HB
    VANSCHADEWIJK, A
    VANGELDERENBOELE, S
    TANKE, HJ
    [J]. CYTOMETRY, 1990, 11 (07): : 784 - 792
  • [7] FLUORESCENCE DECAY TIMES - PROTEINS COENZYMES AND OTHER COMPOUNDS IN WATER
    CHEN, RF
    VUREK, GG
    ALEXANDE.N
    [J]. SCIENCE, 1967, 156 (3777) : 949 - &
  • [8] ENUMERATION OF CR-1 COMPLEMENT RECEPTORS ON ERYTHROCYTES USING A NEW METHOD FOR DETECTING LOW-DENSITY CELL-SURFACE ANTIGENS BY FLOW-CYTOMETRY
    COHEN, JHM
    AUBRY, JP
    JOUVIN, MH
    WIJDENES, J
    BANCHERAU, J
    KAZATCHKINE, M
    REVILLARD, JP
    [J]. JOURNAL OF IMMUNOLOGICAL METHODS, 1987, 99 (01) : 53 - 58
  • [9] TIME-RESOLVED FLOW-CYTOMETRY FOR THE MEASUREMENT OF LANTHANIDE CHELATE FLUORESCENCE .2. INSTRUMENT DESIGN AND EXPERIMENTAL RESULTS
    CONDRAU, MA
    SCHWENDENER, RA
    ZIMMERMANN, M
    MUSER, MH
    GRAF, U
    NIEDERER, P
    ANLIKER, M
    [J]. CYTOMETRY, 1994, 16 (03): : 195 - 205
  • [10] CORRECTION OF CELLULAR AUTOFLUORESCENCE IN FLOW-CYTOMETRY BY MATHEMATICAL-MODELING OF CELLULAR FLUORESCENCE
    CORSETTI, JP
    SOTIRCHOS, SV
    COX, C
    COWLES, JW
    LEARY, JF
    BLUMBURG, N
    [J]. CYTOMETRY, 1988, 9 (06): : 539 - 547