A simple and reliable method to calibrate respiratory magnetometers and Respitrace

被引:86
作者
Banzett, RB
Mahan, ST
Garner, DM
Brughera, A
Loring, SH
机构
[1] Physiology Program, Harvard School of Public Health, Boston, MA 02115
关键词
spirometry; respiratory inductance plethysmography; chest wall; respiratory mechanics; tidal volume; breathing;
D O I
10.1152/jappl.1995.79.6.2169
中图分类号
Q4 [生理学];
学科分类号
071003 ;
摘要
We present a simple and reliable method to calibrate respiratory magnetometers and Respitrace to infer respiratory volume changes. As in earlier methods, we assume two degrees of freedom in the chest wall and that volume displacement depends linearly on surface motion at the rib cage and abdomen. Because the area of the rib cage is larger, a given motion of its surface produces a greater lung volume change; therefore, the rib cage motion signal is given a larger gain before the two signals are added to estimate volume. In contrast to earlier methods, we use a ''standard ratio'' to weight relative gains of the rib cage and abdominal signals for all subjects rather than determining a gain ratio for each individual subject. Our procedure does not require subjects to perform the sometimes difficult isovolume maneuvers used in the calibration method of Konno and Mead (J. Appl. Physiol. 22: 407-422, 1967), does not require statistical computation used in the multiple-breath linear regression method, and does not produce the occasional substantial errors in gain ratio that may occur with the other methods. When magnetometers are used, the standard ratio is 4:1 (rib cage-to-abdomen); when Respitrace is used, the standard ratio is 2:1. In 11 subjects, calibration with standard ratios was as accurate as the isovolume and linear regression techniques. Accuracy during normal breathing was nearly always within 10% (median 2%), but occasional large errors occurred with both instruments.
引用
收藏
页码:2169 / 2176
页数:8
相关论文
共 13 条
[1]  
CHADHA TS, 1982, AM REV RESPIR DIS, V125, P644
[2]  
GOLDMAN MD, 1973, J APPL PHYSIOL, V35, P197, DOI 10.1152/jappl.1973.35.2.197
[3]   MEASUREMENT OF SEPARATE VOLUME CHANGES OF RIB CAGE AND ABDOMEN DURING BREATHING [J].
KONNO, K ;
MEAD, J .
JOURNAL OF APPLIED PHYSIOLOGY, 1967, 22 (03) :407-&
[4]   SINGLE-POSITION CALIBRATION OF THE RESPIRATORY INDUCTANCE PLETHYSMOGRAPH [J].
LOVERIDGE, B ;
WEST, P ;
ANTHONISEN, NR ;
KRYGER, MH .
JOURNAL OF APPLIED PHYSIOLOGY, 1983, 55 (03) :1031-1034
[5]   RIB CAGE DISTORTION DURING VOLUNTARY AND INVOLUNTARY BREATHING ACTS [J].
MCCOOL, FD ;
LORING, SH ;
MEAD, J .
JOURNAL OF APPLIED PHYSIOLOGY, 1985, 58 (05) :1703-1712
[6]   PULMONARY VENTILATION MEASURED FROM BODY SURFACE MOVEMENTS [J].
MEAD, J ;
PETERSON, N ;
GRIMBY, G ;
MEAD, J .
SCIENCE, 1967, 156 (3780) :1383-&
[7]   COMPARISON OF 2-MAGNETOMETER AND 4-MAGNETOMETER METHODS OF MEASURING VENTILATION [J].
ROBERTSON, CH ;
BRADLEY, ME ;
HOMER, LD .
JOURNAL OF APPLIED PHYSIOLOGY, 1980, 49 (03) :355-362
[8]  
SACKNER JD, 1980, AM REV RESPIR DIS, V122, P867
[9]   CALIBRATION OF RESPIRATORY INDUCTIVE PLETHYSMOGRAPH DURING NATURAL BREATHING [J].
SACKNER, MA ;
WATSON, H ;
BELSITO, AS ;
FEINERMAN, D ;
SUAREZ, M ;
GONZALEZ, G ;
BIZOUSKY, F ;
KRIEGER, B .
JOURNAL OF APPLIED PHYSIOLOGY, 1989, 66 (01) :410-420
[10]   COMPUTER-AIDED MEASUREMENT OF BREATH VOLUME AND TIME COMPONENTS USING MAGNETOMETERS [J].
STAGG, D ;
GOLDMAN, M ;
DAVIS, JN .
JOURNAL OF APPLIED PHYSIOLOGY, 1978, 44 (04) :623-633