IMPROVED CONTRAST AT 1.5 TESLA USING HALF-FOURIER IMAGING - APPLICATION TO SPIN-ECHO AND ANGIOGRAPHIC IMAGING

被引:23
作者
HAACKE, EM
MITCHELL, J
LEE, DH
机构
[1] Department of Radiology, Case Western Reserve University, Cleveland
[2] Department of Physics, Case Western Reserve University, Cleveland
[3] General Electric Medical Systems, Milwaukee
[4] School of Medicine, Case Western Reserve University, Cleveland
关键词
Angiographic imaging; Image contrast; Image reconstruction; Short echo times;
D O I
10.1016/0730-725X(90)90216-O
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Half-Fourier imaging is useful for reducing imaging time by requiring less than the usual number of phase-encoding steps. This increase in speed can be traded off for longer repeat times, TR, for improved contrast-to-noise in the same imaging time or to collect short asymmetric echoes. Consequently, it is shown to be especially useful for long TR spin-echo imaging where at 1.5 T a repeat time of 4 sec is recommended for a double-echo TE = 30 90 sequence or 3 sec for a double-echo TE = 15 90 sequence. Short TR FLASH imaging also benefits from a longer TR since there is more time to spoil the signal. In both cases, there is the advantage when a multislice acquisition mode is used that more slices (and hence, a larger volume) can be taken. Another application is to apply half-Fourier imaging in the read direction to avoid spin dephasing and motion artifacts. This is particularly useful in angiographic imaging where smaller pixel sizes and shorter echo times both reduce pixel dephasing. Again, even though taking less than the usual number of data points leads to a reduction in S N, the improved signal and resolution for blood vessels can more than compensate this loss. © 1990.
引用
收藏
页码:79 / 90
页数:12
相关论文
共 24 条
[1]  
Crooks, Arakawa, Hale, Et al., Echo planar imaging at 0.35T with one half of a data set, 5th Annual SMRM, (1986)
[2]  
Cuppen, van Est, Reducing MR imaging time by one-sided reconstruction, Topical conference on fast MRI, (1987)
[3]  
DeMeester, Patrick, Phase corrections for a reconstruction method using data reflection, 5th Annual SMRM, pp. 267-268, (1986)
[4]  
DeMeester, Bearden, Lampman, Holland, The use of conjugate symmetry with 3D volume acquisition, 6th Annual SMRI, (1988)
[5]  
DeMeester, Matched bandwidth asymmetric multiecho sequences, 7th Annual SMRM, (1988)
[6]  
Ehrhardt, Tian, Chang, Reduced MR acquisition time by conjugate reconstruction with off-center order sampling, 7th Anual SMRM, (1988)
[7]  
Feinberg, Hale, Watts, Et al., Halving MR imaging time by conjugation
[8]  
demonstration at 3.5kG, Radiology, 161, pp. 527-531, (1986)
[9]  
Fischer, Schmitt, Barfuss, Bruder, Half-Fourier echo planar imaging at 2 Tesla, 7th Annual SMRM, (1988)
[10]  
Haacke, Bearden, Clayton, Linga, Reduction of MR imaging time by the hybrid fast scan technique, Radiology, 158, pp. 521-529, (1986)