LOW BLANK RHENIUM ISOTOPE RATIO DETERMINATIONS BY V2O5 COATED NICKEL FILAMENTS USING NEGATIVE THERMAL IONIZATION MASS-SPECTROMETRY

被引:20
作者
WALCZYK, T [1 ]
HEBEDA, EH [1 ]
HEUMANN, KG [1 ]
机构
[1] UNIV REGENSBURG,INST ANORGAN CHEM,D-93040 REGENSBURG,GERMANY
来源
INTERNATIONAL JOURNAL OF MASS SPECTROMETRY AND ION PROCESSES | 1994年 / 130卷 / 03期
关键词
RHENIUM; ISOTOPE RATIO; NEGATIVE THERMAL IONIZATION MASS SPECTROMETRY; NICKEL FILAMENTS;
D O I
10.1016/0168-1176(94)85016-X
中图分类号
O64 [物理化学(理论化学)、化学物理学]; O56 [分子物理学、原子物理学];
学科分类号
070203 ; 070304 ; 081704 ; 1406 ;
摘要
Rhenium isotope ratio determinations are, in principle, possible by negative thermal ionization mass spectrometry (NTI-MS). Relatively high rhenium blanks from the commonly-used filament materials prevent accurate isotope ratio determinations, especially for small rhenium sample amounts which are of importance, for example, in geochronology in connection with the Re/Os dating method. Platinum and nickel filaments were tested by different preparation techniques to reduce the rhenium blank contribution from the filament material. The lowest rhenium blank of less than 1 pg was achieved by coating nickel filaments with V2O5 prior to degassing under high vacuum conditions at 850 degrees C. Obviously, the vanadium-nickel oxide layer formed on the surface of the filament during this process prevents further emission of rhenium ions from the filament material. Using Ba(OH)(2) for the enhancement of negative thermal ions, 1 ng of rhenium resulted in ion currents at the detector side of about 10(-11) A with an ionization efficiency of up to 20%. The Re-185/Re-187 isotope ratio of a sample of natural isotopic composition could be determined to be 0.59818 +/- 0.00026 with a relative precision of 0.04%. The isotope ratio determination for an Re-187 spike was comparable in precision but the relative standard deviation of an Re-185 spike was significantly higher, which could be explained by mass fractionations of oxygen in the measured ReO4- ion. The ReO4- ion is about 200 to 2500 times more abundant than the only other detectable rhenium ion in NTI-MSReO(3)(-). The ReO4-/ReO3- ratio decreases with increasing temperature. By the low blank NTI-MS technique described in this work, more precise and accurate determinations of the rhenium isotope ratio and the rhenium concentration by isotope dilution analysis from nanogramme samples are possible.
引用
收藏
页码:237 / 246
页数:10
相关论文
共 19 条
  • [1] RHENIUM IN SEAWATER - CONFIRMATION OF GENERALLY CONSERVATIVE BEHAVIOR
    ANBAR, AD
    CREASER, RA
    PAPANASTASSIOU, DA
    WASSERBURG, GJ
    [J]. GEOCHIMICA ET COSMOCHIMICA ACTA, 1992, 56 (11) : 4099 - 4103
  • [2] COMPARISON OF STABLE ISOTOPE REFERENCE SAMPLES
    COPLEN, TB
    KENDALL, C
    HOPPLE, J
    [J]. NATURE, 1983, 302 (5905) : 236 - 238
  • [3] NEGATIVE THERMAL ION MASS-SPECTROMETRY OF OSMIUM, RHENIUM, AND IRIDIUM
    CREASER, RA
    PAPANASTASSIOU, DA
    WASSERBURG, GJ
    [J]. GEOCHIMICA ET COSMOCHIMICA ACTA, 1991, 55 (01) : 397 - 401
  • [4] ABSOLUTE ISOTOPIC ABUNDANCE RATIO AND ATOMIC WEIGHT OF A REFERENCE SAMPLE OF RHENIUM
    GRAMLICH, JW
    MURPHY, TJ
    GARNER, EL
    SHIELDS, WR
    [J]. JOURNAL OF RESEARCH OF THE NATIONAL BUREAU OF STANDARDS SECTION A-PHYSICS AND CHEMISTRY, 1973, A 77 (06): : 691 - 698
  • [5] GROSSER R, 1986, ADV MASS SPECTROM, P1057
  • [6] HERR W, 1955, Z NATURFORSCH PT A, V10, P613
  • [7] HALF-LIFE OF RHENIUM
    HERR, W
    HINTENBERGER, H
    VOSHAGE, H
    [J]. PHYSICAL REVIEW, 1954, 95 (06): : 1691 - 1691
  • [8] Heumann K.G., 1988, INORGANIC MASS SPECT, P301
  • [9] APPLICATION OF AN ECONOMICAL AND SMALL THERMAL IONIZATION MASS-SPECTROMETER FOR ACCURATE ANION TRACE ANALYSES
    HEUMANN, KG
    SCHINDLMEIER, W
    ZEININGER, H
    SCHMIDT, M
    [J]. FRESENIUS ZEITSCHRIFT FUR ANALYTISCHE CHEMIE, 1985, 320 (05): : 457 - 462
  • [10] RADIOGENIC OSMIUM FROMRHENIUM-CONTAINING MOLYBDENITE
    HINTENBERGER, H
    HERR, W
    VOSHAGE, H
    [J]. PHYSICAL REVIEW, 1954, 95 (06): : 1690 - 1691