PRELIMINARY-RESULTS FROM ATTENUATION-SLOPE MAPPING OF PLAQUE USING INTRAVASCULAR ULTRASOUND

被引:56
作者
WILSON, LS [1 ]
NEALE, ML [1 ]
TALHAMI, HE [1 ]
APPLEBERG, M [1 ]
机构
[1] ROYAL N SHORE HOSP,DEPT VASC SURG,SYDNEY,NSW,AUSTRALIA
关键词
INTRAVASCULAR ULTRASOUND; DEGENERATIVE PLAQUE; TISSUE CHARACTERIZATION; ULTRASONIC ATTENUATION;
D O I
10.1016/0301-5629(94)90089-2
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Excised femoral and iliac artery segments have been examined with 20 MHz intravascular ultrasound followed by histological assessment. During the ultrasound examinations, radio-frequency (RF) data were recorded digitally, and used for calculating local values of attenuation slope throughout the tissue, using a frequency-domain technique. The RF data were also reconstructed as conventional ultrasound images, and the attenuation-slope information presented as a thresholded colour overlay. Areas identified as degenerative plaque in the histological assessments were usually found to correspond to areas of high attenuation slope, and were clearly identified from the pattern of colours on the combined image. Some examples are presented, illustrating the appearance of various pathologies imaged by this technique.
引用
收藏
页码:529 / 542
页数:14
相关论文
共 13 条
[1]   ANGLE-DEPENDENT BACKSCATTER FROM THE ARTERIAL-WALL [J].
DEKROON, MGM ;
VANDERWAL, LF ;
GUSSENHOVEN, WJ ;
BOM, N .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1991, 17 (02) :121-126
[2]   CONTRIBUTION OF LOCALIZED CALCIUM DEPOSITS TO DISSECTION AFTER ANGIOPLASTY - AN OBSERVATIONAL STUDY USING INTRAVASCULAR ULTRASOUND [J].
FITZGERALD, PJ ;
PORTS, TA ;
YOCK, PG .
CIRCULATION, 1992, 86 (01) :64-70
[3]   AN APPROACH TO DIGITAL SIGNAL-PROCESSING FOR ULTRASONICS RESEARCH [J].
KNIGHT, PC ;
ROBINSON, DE .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1987, 13 (06) :345-352
[4]   ESTIMATING THE ACOUSTIC ATTENUATION COEFFICIENT SLOPE FOR LIVER FROM REFLECTED ULTRASOUND SIGNALS [J].
KUC, R ;
SCHWARTZ, M .
IEEE TRANSACTIONS ON SONICS AND ULTRASONICS, 1979, 26 (05) :353-362
[5]   STRUCTURE-DEPENDENT DYNAMIC MECHANICAL-BEHAVIOR OF FIBROUS CAPS FROM HUMAN ATHEROSCLEROTIC PLAQUES [J].
LEE, RT ;
GRODZINSKY, AJ ;
FRANK, EH ;
KAMM, RD ;
SCHOEN, FJ .
CIRCULATION, 1991, 83 (05) :1764-1770
[6]  
Linker D T, 1991, Int J Card Imaging, V6, P255, DOI 10.1007/BF01797857
[7]   MEASUREMENT OF THE ULTRASONIC PROPERTIES OF VASCULAR TISSUES AND BLOOD FROM 35-65 MHZ [J].
LOCKWOOD, GR ;
RYAN, LK ;
HUNT, JW ;
FOSTER, FS .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1991, 17 (07) :653-666
[8]   EFFECTS OF FIBROUS CAP THICKNESS ON PEAK CIRCUMFERENTIAL STRESS IN MODEL ATHEROSCLEROTIC VESSELS [J].
LOREE, HM ;
KAMM, RD ;
STRINGFELLOW, RG ;
LEE, RT .
CIRCULATION RESEARCH, 1992, 71 (04) :850-858
[9]   ELIMINATION OF DIFFRACTION ERROR IN ACOUSTIC ATTENUATION ESTIMATION VIA AXIAL BEAM TRANSLATION [J].
OPHIR, J ;
MEHTA, D .
ULTRASONIC IMAGING, 1988, 10 (02) :139-152
[10]   BEAM PATTERN (DIFFRACTION) CORRECTION FOR ULTRASONIC-ATTENUATION MEASUREMENT [J].
ROBINSON, DE ;
WILSON, LS ;
BIANCHI, T .
ULTRASONIC IMAGING, 1984, 6 (03) :293-303