A DIRECT-READING DEVICE FOR MEASUREMENT OF PATCH-CLAMP MICROPIPETTE TIP DIAMETERS

被引:13
作者
MARTIN, DK [1 ]
COOK, DI [1 ]
机构
[1] UNIV SYDNEY,DEPT PHYSIOL,SYDNEY,NSW 2006,AUSTRALIA
来源
PFLUGERS ARCHIV-EUROPEAN JOURNAL OF PHYSIOLOGY | 1990年 / 417卷 / 03期
关键词
Electronic device; Inner tip diameter; Micropipette; Patch clamp;
D O I
10.1007/BF00370989
中图分类号
Q4 [生理学];
学科分类号
071003 ;
摘要
This paper describes an electronic device for the measurement of the diameter of the tip opening of micropipettes. The device measures the threshold pressure required to produce bubbles from the micropipette submerged in a liquid and displays the tip inner diameter in micrometres. We demonstrate that this technique can be used effectively in a range of liquids of different density, surface tension and viscosity and that the thickness of the glass from which the micropipettes are made does not influence the measurements. © 1990 Springer-Verlag.
引用
收藏
页码:255 / 258
页数:4
相关论文
共 11 条
[1]   NEW MICRO-ELECTRODE TECHNIQUES FOR INTRACELLULAR WORK IN SMALL CELLS [J].
BROWN, KT ;
FLAMING, DG .
NEUROSCIENCE, 1977, 2 (06) :813-827
[2]  
Clift R., 1978, BUBBLES DROPS PARTIC, P321
[3]  
COREY DP, 1983, SINGLE CHANNEL RECOR, P66
[4]  
Davidson J., 1960, T I CHEM ENGRS, V38, P335
[5]  
KUPFERBERG A, 1969, TRANS INST CHEM ENG, V47, pT241
[6]  
Landau L. D., 1976, FLUID MECH-SOV RES, V1
[7]   BUBBLE PRESSURE MEASUREMENT OF MICROPIPET TIP OUTER DIAMETER [J].
MITTMAN, S ;
FLAMING, DG ;
COPENHAGEN, DR ;
BELGUM, JH .
JOURNAL OF NEUROSCIENCE METHODS, 1987, 22 (02) :161-166
[8]  
PURVES RD, 1981, MICROELECTRODE METHO, P146
[9]   RATES OF DIFFUSIONAL EXCHANGE BETWEEN SMALL CELLS AND A MEASURING PATCH PIPETTE [J].
PUSCH, M ;
NEHER, E .
PFLUGERS ARCHIV-EUROPEAN JOURNAL OF PHYSIOLOGY, 1988, 411 (02) :204-211
[10]   FORMATION OF GAS BUBBLES AT NOZZLES AT CONSTANT VOLUME THROUGHPUT [J].
RUFF, K .