COMPARISON OF A TYMPANIC THERMOMETER TO RECTAL AND ORAL THERMOMETERS IN A PEDIATRIC EMERGENCY DEPARTMENT

被引:49
作者
CHAMBERLAIN, JM
GRANDNER, J
RUBINOFF, JL
KLEIN, BL
WAISMAN, Y
HUEY, M
机构
关键词
D O I
10.1177/0009922891030004S08
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
We performed a clinical trial of a new tympanic thermometer to test its accuracy in a pediatric emergency department. Tympanic temperature was compared to oral or rectal glass and electronic temperature, depending on the developmental age of the child. Results were controlled for age of the child, cooperation, quantity of cerumen, and the presence or absence of otitis media. Our results suggest good correlation of tympanic with oral and rectal glass thermometry except in infants less than three months of age. Sensitivity and specificity, respectively, were 80% and 93% for detecting fever of 38-degrees-C and 80% and 95% for detecting fever of 38.5-degrees-C.
引用
收藏
页码:24 / 29
页数:6
相关论文
共 14 条
[1]  
Barber N, 1989, Pediatr Nurs, V15, P477
[2]  
BECK WC, 1975, GUTHRIE B, V44, P175
[3]  
Benzinger M, 1972, 5TH NBS S TEMP PITTS, P2089
[4]  
ERICKSON R, 1980, NURS RES, V29, P157
[5]  
GERBRANDY J, 1954, CLIN SCI, V13, P615
[6]  
KENNEY RD, 1990, PEDIATRICS, V85, P854
[7]  
LOBBES JM, THESIS U FLORIDA
[8]   COMPARISON OF RECTAL, AXILLARY, AND TYMPANIC MEMBRANE TEMPERATURES IN INFANTS AND YOUNG-CHILDREN [J].
MUMA, BK ;
TRELOAR, DJ ;
WURMLINGER, K ;
PETERSON, E ;
VITAE, A .
ANNALS OF EMERGENCY MEDICINE, 1991, 20 (01) :41-44
[9]   ASSESSMENT OF AN AURAL INFRARED-SENSOR FOR BODY-TEMPERATURE MEASUREMENT IN CHILDREN [J].
RHOADS, FA ;
GRANDNER, J .
CLINICAL PEDIATRICS, 1990, 29 (02) :112-115