IODINE LOSSES DURING WINKLER TITRATIONS

被引:8
作者
KNAPP, GP
STALCUP, MC
STANLEY, RJ
机构
[1] Woods Hole Oceanographic Institution, Woods Hole
来源
DEEP-SEA RESEARCH PART A-OCEANOGRAPHIC RESEARCH PAPERS | 1991年 / 38卷 / 01期
基金
美国国家科学基金会;
关键词
D O I
10.1016/0198-0149(91)90057-M
中图分类号
P7 [海洋学];
学科分类号
0707 ;
摘要
An experiment designed to measure iodine loss during the aliquot version of the Winkler titration for dissolved oxygen in seawater shows that 0.01-0.03 ml l-1 equivalent oxygen is lost at typical oceanic concentrations in the method presently used. A standardization technique, which mimics that employed during the titration of seawater samples, compensates for this iodine loss throughout the oceanic range. This result, contradicting an earlier report by GREEN and CARRITT (1966, Analyst, 91, 207-208), demonstrates that the whole-bottle method of oxygen titration is not to be preferred over the aliquot method.
引用
收藏
页码:121 / 128
页数:8
相关论文
共 8 条
[1]  
CARPENTER JAMES H., 1965, LIMNOL OCEANOGR, V10, P135
[2]  
CARRITT DE, 1966, J MAR RES, V24, P286
[3]   AUTOMATED AMPEROMETRIC OXYGEN TITRATION [J].
CULBERSON, CH ;
HUANG, S .
DEEP-SEA RESEARCH PART A-OCEANOGRAPHIC RESEARCH PAPERS, 1987, 34 (5-6) :875-880
[4]   AN IMPROVED IODINE DETERMINATION FLASK FOR WHOLE-BOTTLE TITRATIONS [J].
GREEN, EJ ;
CARRITT, DE .
ANALYST, 1966, 91 (1080) :207-&
[5]  
KNAPP GP, 1989, WHOI8923 WOODS HOL O
[6]  
SAUNDERS PM, 1986, J PHYS OCEANOGR, V16, P189, DOI 10.1175/1520-0485(1986)016<0189:TAOMOS>2.0.CO
[7]  
2
[8]  
Strickland J.D.H., 1972, PRACTICAL HDB SEAWAT, V2nd, DOI [10.25607/OBP-1791, DOI 10.25607/OBP-1791]