AN INTERVAL-TO-FREQUENCY CONVERTER FOR NEUROPHYSIOLOGICAL USE

被引:4
作者
FRANZ, GN
SPELMAN, FA
机构
[1] Department of Physiology and Biophysics, Regional Primate Research Center, University of Washington, Seattle
来源
ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY | 1968年 / 25卷 / 06期
关键词
D O I
10.1016/0013-4694(68)90240-X
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
An interval-to-frequency converter is described. Design is simplified by use of a varistor, instead of a complicated diode network, to create a hyperbolic capacitor discharge. The capacitor voltage is sampled with a field-effect transistor at the end of the interval between two succeeding spikes in order to give an analog representation of the reciprocal of the interval. © 1968.
引用
收藏
页码:582 / &
相关论文
共 2 条
[1]  
KAY RH, 1965, ELECTRON ENG, V37, P543
[2]  
LAMB WGP, 1965, ELECTRON ENG, V37, P809