SAMPLING METHODS FOR A MATRIX IONIZATION-CHAMBER SYSTEM

被引:11
作者
VANHERK, M
BIJHOLD, J
HOOGERVORST, B
MEERTENS, H
机构
[1] Radiotherapy Department, Netherlands Cancer Institute (Antoni van Leeuwenhoek Huis), 1066, CX, Amsterdam
关键词
IONIZATION CHAMBER; PORTAL IMAGING; CODED SAMPLING; WEIGHING DESIGNS;
D O I
10.1118/1.596829
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
To achieve increased image acquisition speed or better image quality, several read-out methods for a matrix ionization chamber system have been investigated. In this device, which is applied for portal imaging in radiation therapy, 256 x 256 small liquid-filled ionization chambers are scanned by switching the polarizing voltage applied to rows of chambers. The ionization current of each column is measured by a separate amplifier. In this approach, instead of measuring row by row, more complex switching schemes can be applied for the polarizing voltage. These schemes are useful either for varying speed and spatial resolution of the imaging device or for coded sampling of the ionization signal. The former option allows for doubling or quadrupling the acquisition speed with a small loss in image quality, or for obtaining a large improvement in signal-to-noise ratio at the cost of image resolution. In the latter option, coded sampling, the image is reconstructed mathematically from the measured signals. It is shown that in this case the application of Hadamard or derived matrices for sampling leads, under certain circumstances, to a noise reduction in the reconstructed image.
引用
收藏
页码:409 / 418
页数:10
相关论文
共 13 条
  • [1] RADIATION-FIELD EDGE-DETECTION IN PORTAL IMAGES
    BIJHOLD, J
    GILHUIJS, KGA
    VANHERK, M
    MEERTENS, H
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1991, 36 (12) : 1705 - 1710
  • [2] Gonzalez R. C., 1977, DIGITAL IMAGE PROCES
  • [3] SPECTROMETRIC IMAGER
    HARWIT, M
    [J]. APPLIED OPTICS, 1971, 10 (06): : 1415 - &
  • [4] Some improvements in weighing and other experimental techniques
    Hotelling, H
    [J]. ANNALS OF MATHEMATICAL STATISTICS, 1944, 15 : 297 - 306
  • [5] A TEST OBJECT FOR EVALUATION OF PORTAL FILMS
    LUTZ, WR
    BJARNGARD, BE
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1985, 11 (03): : 631 - 634
  • [6] 1ST CLINICAL-EXPERIENCE WITH A NEWLY DEVELOPED ELECTRONIC PORTAL IMAGING DEVICE
    MEERTENS, H
    VANHERK, M
    BIJHOLD, J
    BARTELINK, H
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1990, 18 (05): : 1173 - 1181
  • [7] AN INVERSE FILTER FOR DIGITAL RESTORATION OF PORTAL IMAGES
    MEERTENS, H
    VANHERK, M
    WEEDA, J
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1988, 33 (06) : 687 - 702
  • [8] ON HOTELLINGS WEIGHING PROBLEM
    MOOD, AM
    [J]. ANNALS OF MATHEMATICAL STATISTICS, 1946, 17 (04): : 432 - 446
  • [9] Raghavarao, 1971, CONSTRUCTIONS COMBIN
  • [10] Stoer J., 1993, INTRO NUMERICAL ANAL