GAS-DIFFUSION WITH PRECONCENTRATION FOR THE DETERMINATION OF FLUORIDE IN WATER SAMPLES BY FLOW-INJECTION

被引:15
作者
CARDWELL, TJ
CATTRALL, RW
MITRI, M
机构
[1] Centre for Scientific Instrumentation, Chemistry Department, La Trobe University, Bundoora
关键词
D O I
10.1016/0039-9140(94)80176-2
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The preconcentration of fluoride is achieved on-line by converting it to trimethylsilane which then diffuses through a gas permeable membrane to be absorbed in a stationary sodium hydroxide acceptor stream. This stream is enclosed in the sample loop of an injection valve and after preconcentration, the fluoride sample is flushed into a flow injection manifold for spectrophotometric analysis by the zirconium/alizarin S procedure at 520 nm. The method is suitable for fluoride analysis in the range 0.1-10 mg/l at a sampling rate of 17/hr. Phosphate does not interfere and aluminium and iron can be tolerated at 200 and 500 times the fluoride concentration, respectively. The LOD was calculated to be 0.055 mg/l and LOQ was found to be 0.18 mg/l.
引用
收藏
页码:115 / 123
页数:9
相关论文
共 18 条
[1]   SOLUTE TRANSFER IN ONLINE ANALYTICAL FLOW-THROUGH DIALYZERS [J].
BERNHARDSSON, B ;
MARTINS, E ;
JOHANSSON, G .
ANALYTICA CHIMICA ACTA, 1985, 167 (JAN) :111-122
[2]   OPTIMIZATION OF PARAMETERS FOR GAS-DIFFUSION FLOW-INJECTION SYSTEMS [J].
CANHAM, JS ;
GORDON, G ;
PACEY, GE .
ANALYTICA CHIMICA ACTA, 1988, 209 (1-2) :157-163
[3]   FLOW-INJECTION SPECTROPHOTOMETRIC DETERMINATION OF FLUORIDE BY USING THE ZIRCONIUM ALIZARIN RED-S COMPLEX [J].
CARDWELL, TJ ;
CATTRALL, RW ;
MITRI, M ;
HAMILTON, IC .
ANALYTICA CHIMICA ACTA, 1988, 214 (1-2) :433-438
[4]   IN-STREAM MASKING OF ALUMINUM IN THE DETERMINATION OF FLUORIDE BY FLOW-INJECTION POTENTIOMETRY [J].
DAVEY, DE ;
MULCAHY, DE ;
MUGGLETON, TJ ;
OCONNELL, GR .
ANALYTICAL LETTERS, 1992, 25 (03) :607-624
[5]  
DINGLI G, 1987, ANAL P, V24, P342
[6]  
FANG ZL, 1988, ANAL CHIM ACTA, V214, P41
[7]  
MACDONALD AMG, 1983, 9TH INT S MICR TECHN
[8]   A MODIFIED ZIRCONIUM-ALIZARIN METHOD FOR DETERMINING FLUORIDE IN NATURAL WATERS [J].
MEYLING, AH ;
MEYLING, J .
ANALYST, 1963, 88 (104) :84-&
[9]   ESTIMATION OF FLUORINE IN MICROGRAMME QUANTITIES A MODIFIED PROCEDURE [J].
MURTY, GVLN ;
VISWANATHAN, TS ;
RAMAKRISHNA, V .
ANALYTICA CHIMICA ACTA, 1957, 16 (03) :213-215
[10]   DETERMINATION OF AMMONIUM ION IN A FLOW-INJECTION SYSTEM WITH A GAS-DIFFUSION MEMBRANE - SELECTION OF OPTIMAL CONDITIONS FOR THE PH INDICATOR [J].
NAKATA, R ;
KAWAMURA, T ;
SAKASHITA, H ;
NITTA, A .
ANALYTICA CHIMICA ACTA, 1988, 208 (1-2) :81-90