PHASE-CONTRAST IMAGING-SYSTEM FOR TEXT-U

被引:10
作者
CHATTERJEE, R
HALLOCK, GA
GARTMAN, ML
机构
[1] University of Texas at Austin, Austin
关键词
D O I
10.1063/1.1146380
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A diagnostic to study plasma density fluctuations, phase contrast imaging has been developed for the Texas Experimental Tokamak-Upgrade. The diagnostic has a sensitivity of about 10-4 ne0 and is capable of detecting a wide range of wave numbers (0.5-12 cm-1) with a bandwidth of 500 kHz. The design of the diagnostic, some results of acoustic calibration tests, and preliminary results of simulation of expected spectra are presented. © 1995 American Institute of Physics.
引用
收藏
页码:457 / 459
页数:3
相关论文
共 7 条
[1]  
BRAVENEC RV, 1992, PHYS FLUIDS B-PLASMA, V7, P2127
[2]   A PHASE-CONTRAST INTERFEROMETER ON DIII-D [J].
CODA, S ;
PORKOLAB, M ;
CARLSTROM, TN .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1992, 63 (10) :4974-4976
[3]   THE TEXT UPGRADE VERTICAL INTERFEROMETER [J].
HALLOCK, GA ;
GARTMAN, ML ;
LI, W ;
CHIANG, K ;
SHIN, S ;
CASTLES, RL ;
CHATTERJEE, R ;
RAHMAN, AS .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1992, 63 (10) :4987-4989
[4]  
Korpel A, 1988, ACOUSTO OPTICS
[5]   APPLICABILITY OF LASER PHASE-CONTRAST METHOD FOR THE MEASUREMENTS OF ELECTRON-DENSITY FLUCTUATIONS IN HIGH-TEMPERATURE PLASMAS [J].
TANAKA, K ;
MATSUO, K ;
GOTO, K ;
BOWDEN, M ;
MURAOKA, K ;
FURUKAWA, T ;
SUDO, S ;
OBIKI, T .
JAPANESE JOURNAL OF APPLIED PHYSICS PART 1-REGULAR PAPERS SHORT NOTES & REVIEW PAPERS, 1992, 31 (07) :2260-2262
[6]   IMAGING METHODS FOR THE OBSERVATION OF PLASMA-DENSITY FLUCTUATIONS [J].
WEISEN, H .
PLASMA PHYSICS AND CONTROLLED FUSION, 1986, 28 (08) :1147-1159
[7]  
ZERNIKE F, 1942, PHYSICA UTRECHT, V7, P686