PACED BEATS FOLLOWING SINGLE NONSENSED COMPLEXES IN A CODEPENDENT CARDIOVERTER DEFIBRILLATOR AND BRADYCARDIA PACING SYSTEM - POTENTIAL FOR VENTRICULAR-TACHYCARDIA INDUCTION

被引:20
作者
CALLANS, DJ
HOOK, BG
MARCHLINSKI, FE
机构
[1] HOSP UNIV PENN,CLIN ELECTROPHYSIOL LAB,3400 SPRUCE ST,9 FOUNDERS PAVIL,PHILADELPHIA,PA 19104
[2] UNIV PENN,SCH MED,DEPT MED,CARDIOVASC SECT,PHILADELPHIA,PA 19104
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1991年 / 14卷 / 08期
关键词
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR; CARDIAC PACING; SENSING ALGORITHMS; VENTRICULAR ARRHYTHMIAS;
D O I
10.1111/j.1540-8159.1991.tb02868.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Patients with implantable defibrillators often require bradycardia pacemakers. Adverse interactions between separate defibrillator and bradycardia pacing units have occurred, including failure to detect ventricular fibrillation due to persistent bradycardia pacing during the arrhythmia. A device with combined bradycardia pacing and antitachycardia therapy capability may obviate adverse device interactions. We describe a previously unrecognized phenomenon that may occur in a combined device when the algorithms for sensing bradycardia and tachycardia are "codependent"; that is, the circuitry for brady- and tachyarrhythmia detection relies on the same automatic gain sense amplifier. Three of 37 patients in whom the device was implanted had ventricular tachycardia initiated when bradycardia pacing stimuli were delivered by the device after probable nonsensed sinus beats. In each case, nonsensed beats appeared to have a markedly diminished amplitude, occurred after ventricular premature depolarizations that produced large amplitude electrograms, and had an electrogram morphology that matched that of sinus rhythm. In each case, the bradycardia pacing interval was at least 1,200 msec (range 1,200 to 1,714 msec). In two of the three patients, large amplitude ventricular premature depolarizations or nonsustained ventricular tachycardia caused an adjustment of the gain control that potentiated the failure to sense the subsequent lower amplitude signal. In all three patients, the induced arrhythmia was rapidly terminated by pacing or cardioversion. Decreasing the bradycardia pacing interval by 110-514 msec has prevented recurrence during short-term follow-up. Our findings suggest that codependent bradycardia and antitachycardia devices may have their own unique potential difficulties in adapting to rapid changes in rate and signal amplitude.
引用
收藏
页码:1281 / 1287
页数:7
相关论文
共 9 条
  • [1] CLINICAL INTERACTIONS BETWEEN PACEMAKERS AND AUTOMATIC IMPLANTABLE CARDIOVERTER-DEFIBRILLATORS
    CALKINS, H
    BRINKER, J
    VELTRI, EP
    GUARNIERI, T
    LEVINE, JH
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1990, 16 (03) : 666 - 673
  • [2] CLINICAL-EXPERIENCE, COMPLICATIONS, AND SURVIVAL IN 70 PATIENTS WITH THE AUTOMATIC IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
    ECHT, DS
    ARMSTRONG, K
    SCHMIDT, P
    OYER, PE
    STINSON, EB
    WINKLE, RA
    [J]. CIRCULATION, 1985, 71 (02) : 289 - 296
  • [3] COMBINED AUTOMATIC IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR AND PACEMAKER SYSTEMS - IMPLANTATION TECHNIQUES AND FOLLOW-UP
    EPSTEIN, AE
    KAY, GN
    PLUMB, VJ
    SHEPARD, RB
    KIRKLIN, JK
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1989, 13 (01) : 121 - 131
  • [4] HOOK BG, 1991, J AM COLL CARDIOL, V17, P985
  • [5] UNIPOLAR PACER ARTIFACTS INDUCED FAILURE OF AN AUTOMATIC IMPLANTABLE CARDIOVERTER DEFIBRILLATOR TO DETECT VENTRICULAR-FIBRILLATION
    KIM, SG
    FURMAN, S
    WASPE, LE
    BRODMAN, R
    FISHER, JD
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1986, 57 (10) : 880 - 881
  • [6] COMBINED BIPOLAR DUAL CHAMBER PACING AND AUTOMATIC IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
    RUFFY, R
    LAL, R
    KOUCHOUKOS, NT
    KIM, SS
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1986, 7 (04) : 933 - 937
  • [7] DEVELOPMENTS FOR FUTURE IMPLANTABLE CARDIOVERTERS AND DEFIBRILLATORS
    SAKSENA, S
    LINDSAY, BD
    PARSONNET, V
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1987, 10 (06): : 1342 - 1358
  • [8] SINGER I, 1988, PACE, V11, P485
  • [9] IMPLANTABLE CARDIOVERTERS AND DEFIBRILLATORS
    TROUP, PJ
    [J]. CURRENT PROBLEMS IN CARDIOLOGY, 1989, 14 (12) : 675 - 843