HEIDELBERG RETINA TOMOGRAPH - REPRODUCIBILITY AND RELATIVE ERROR IN THE MODEL EYE AS A FUNCTION OF THE WIDTH OF THE PUPIL

被引:3
作者
JANKNECHT, P
FUNK, J
机构
关键词
D O I
10.1055/s-2008-1045499
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Background We measured the reproducibility and relative error of the Heidelberg Retina Tomograph (HRT) in a model eye with various widths of the pupil. Methods We put diaphragms of various diameters in front of the lens of a model eye. The volume of an artificial papilla which was inserted into the model eye was measured three times. The coefficient of variation (standard deviation/mean value) was taken as a measure of the reproducibility. The relative error was defined as the absolute value of the quotient of the difference between measured and real value. Results The relative error for diaphragms of 1.5 to 5 mm was 7% to 18.4%, for a diaphragm of 1 mm 19.8%, and for a diaphragm of 11 mm 21.3%. The coefficient of variation ranged from 0.49% to 1.95% for diaphragms of 1.5 to 11 mm diameter, and was 8.1 for a diaphragm of 1 mm diameter. Conclusion The HRT is capable of measuring volumes with a reasonable accuracy. The accuracy does not depend on the width of the pupil over a wide range of pupil diameters. In cases of very small or very much dilated pupils the accuracy and reproducibility declines.
引用
收藏
页码:98 / 102
页数:5
相关论文
共 10 条
[1]  
DREHER AW, 1991, INVEST OPHTH VIS SCI, V32, P2992
[2]   REPRODUCIBILITY OF TOPOGRAPHIC MEASUREMENTS OF THE NORMAL AND GLAUCOMATOUS OPTIC-NERVE HEAD WITH THE LASER TOMOGRAPHIC SCANNER [J].
DREHER, AW ;
TSO, PC ;
WEINREB, RN .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 1991, 111 (02) :221-229
[3]  
KLINGBEIL U, 1990, NONINVASIVE DIAGNOST
[4]  
KRUSE FE, 1989, OPHTHALMOLOGY, V96, P1320
[5]  
LUSKY M, 1993, INVEST OPHTH VIS SCI, V34, P1507
[6]   REPRODUCIBILITY OF TOPOMETRIC DATA-ACQUISITION IN NORMAL AND GLAUCOMATOUS OPTIC-NERVE HEADS WITH THE LASER TOMOGRAPHIC SCANNER [J].
ROHRSCHNEIDER, K ;
BURK, ROW ;
VOLCKER, HE .
GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 1993, 231 (08) :457-464
[7]  
SHAFRANOV G, 1993, INVEST OPHTH VIS SCI, V34, P1507
[8]  
THIEL A, 1993, INVEST OPHTH VIS SCI, V34, P1506
[9]   EFFECT OF REPETITIVE IMAGING ON TOPOGRAPHIC MEASUREMENTS OF THE OPTIC-NERVE HEAD [J].
WEINREB, RN ;
LUSKY, M ;
BARTSCH, DU ;
MORSMAN, D .
ARCHIVES OF OPHTHALMOLOGY, 1993, 111 (05) :636-638
[10]  
WEINREB RN, 1990, SCANNING LASER OPHTH