A NEW SPRAY CHAMBER FOR INDUCTIVELY COUPLED PLASMA SPECTROMETRY

被引:27
作者
WU, M [1 ]
HIEFTJE, GM [1 ]
机构
[1] INDIANA UNIV,DEPT CHEM,BLOOMINGTON,IN 47405
关键词
SPRAY CHAMBER; MEMORY EFFECT; FIGURES OF MERIT; INDUCTIVELY COUPLED PLASMA; ATOMIC SPECTROMETRY;
D O I
10.1366/0003702924123647
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A new type of spray chamber is evaluated for inductively coupled plasma emission spectrometry. The new design combines gravitational, centrifugal, turbulent and impact loss mechanisms in one apparatus to efficiently remove large droplets, increase transport efficiency and reduce memory effects. Compared with the commonly used Scott-type spray chamber, the new vertical rotary spray chamber has at least 30% higher sample-utilization efficiency, 2-3 times shorter sample clean-out time., half the cost, and simplified construction. Moreover, it offers somewhat better signal-to-background ratios, detection limits, and precision.
引用
收藏
页码:1912 / 1918
页数:7
相关论文
共 38 条
[1]   DIRECT ATOMIC SPECTROMETRIC ANALYSIS BY SLURRY ATOMIZATION .8. FLOW-INJECTION INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION-SPECTROMETRY [J].
AMBROSE, AJ ;
EBDON, L ;
FOULKES, ME ;
JONES, P .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1989, 4 (02) :219-222
[2]   SOME EFFECTS OF HIGH-SOLIDS MATRICES ON THE SAMPLE DELIVERY SYSTEM AND THE MEINHARD CONCENTRIC NEBULIZER DURING ICP EMISSION ANALYSES [J].
BAGINSKI, BR ;
MEINHARD, JE .
APPLIED SPECTROSCOPY, 1984, 38 (04) :568-572
[3]   PRESSURE MEASUREMENTS IN THE NEBULIZER SPRAY CHAMBER OF AN INDUCTIVITY COUPLED PLASMA [J].
BELCHAMBER, RM ;
HORLICK, G .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1981, 36 (06) :581-583
[4]  
BORER MW, 1992, IN PRESS SPECTROCHIM
[5]   MODIFICATION AND OPTIMIZATION OF A 50 MHZ INDUCTIVELY COUPLED ARGON PLASMA WITH SPECIAL REFERENCE TO ANALYSES USING ORGANIC-SOLVENTS [J].
BOUMANS, PWJM ;
LUXSTEINER, MC .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1982, 37 (02) :97-126
[6]  
Browner R. F., 1987, INDUCTIVELY COUPLE 2, P244
[7]   EVOLUTION OF DROP SIZE DISTRIBUTIONS FOR PNEUMATICALLY GENERATED AEROSOLS IN INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION-SPECTROMETRY [J].
CANALS, A ;
HERNANDIS, V ;
BROWNER, RF .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1990, 45 (06) :591-601
[8]  
CHRISTCHURCH NZ, 1985, CAN J CHEM ENG, V55, P466
[9]  
Davies C.N., 1953, P I MECH ENG, V1, P185, DOI DOI 10.1177/002034835316701B13
[10]   JET-IMPACT NEBULIZATION FOR SAMPLE INTRODUCTION IN INDUCTIVELY COUPLED PLASMA SPECTROMETRY [J].
DOHERTY, MP ;
HIEFTJE, GM .
APPLIED SPECTROSCOPY, 1984, 38 (03) :405-412