EFFECTIVE DOSE AND RISK ASSESSMENT FROM FILM TOMOGRAPHY USED FOR DENTAL IMPLANT DIAGNOSTICS

被引:49
作者
FREDERIKSEN, NL
BENSON, BW
SOKOLOWSKI, TW
机构
[1] Baylor College of Dentistry, Baylor University Medical Center, Dallas, Texas
关键词
RADIOGRAPHY; PANORAMIC; DENTAL; TOMOGRAPHY; X-RAY; RADIATION; DOSAGE;
D O I
10.1259/dmfr.23.3.7835511
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
There are no data relating complex film tomography with effective dose that may be used to estimate the relative risk associated with dental implant diagnostics. The purpose of this study was to calculate the effective dose and estimate risk from the use of the Scanora(R) multimodal imaging system. With the use of a tissue equivalent human phantom and thermoluminescent dosimetry, panoramic radiography was found to result in an effective dose of 26 muSv, while complex film tomography resulted in an effective dose of <1 muSv to 30 muSv depending on the anatomical location of the imaging plane and the collimation option. An effective dose of this magnitude for panoramic radiography was estimated to represent a probability for stochastic effects on the order of 1.9 X 10(-6). Similarly, the effective dose associated with film tomography may be estimated to be equal to a probability for stochastic effects in the range of much less than 1 X 10(-6) to 2.2 X 10(-6).