DOSE BEHIND VARIOUS IMMOBILIZATION AND BEAM-MODIFYING DEVICES

被引:36
作者
MELLENBERG, DE [1 ]
机构
[1] UNIV MICHIGAN,ANN ARBOR,MI 48109
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 1995年 / 32卷 / 04期
关键词
SURFACE DOSE; DOSE BUILDUP; IMMOBILIZATION; COMPENSATORS;
D O I
10.1016/0360-3016(94)00371-Q
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To quantify the degradation of skin sparing associated with using beam modifiers such as compensators, immobilization devices, and custom blocks for high energy photon beams. Methods and Materials: The degradation of skin sparing was quantified by measuring dose build-up curves with an extrapolation chamber for 6 and 15 MV photon beams, Uniform thickness compensators made of gypsum and lead, thermoplastic mask material, immobilization cradle foam, and cerrobend custom blocks were placed in geometries that mimic relevant clinical situations. Results: Compensators, whether made of gypsum or lead, placed in the linear accelerator's wedge slot did not significantly effect the depth dose curve's build-up region, Immobilization devices such as cradle foam or thermoplastic placed in contact with the patient degrade the skin sparing expected from high energy photon beams proportional to their thickness and density, Measurements behind custom blocks show that surface and near surface doses for a blocked field are best described by build-up curves for an equivalent size open field. Conclusions: These results allow explanation and possibly prediction of skin reactions on patients in which compensators, foam immobilization cradles, thermoplastic masks, or custom blocks are used, These results also provide a baseline by which substitute materials can be evaluated.
引用
收藏
页码:1193 / 1197
页数:5
相关论文
共 9 条
[1]   CHARACTERIZATION OF GYPSUM ATTENUATORS FOR RADIOTHERAPY DOSE MODIFICATION [J].
ARORA, VR ;
WEEKS, KJ .
MEDICAL PHYSICS, 1994, 21 (01) :77-81
[2]   DOSE ENHANCEMENT IN BUILDUP REGION BY LEAD, ALUMINUM, AND LUCITE ABSORBERS FOR 15 MVP PHOTON-BEAM [J].
CIESIELSKI, B ;
REINSTEIN, LE ;
WIELOPOLSKI, L ;
MEEK, A .
MEDICAL PHYSICS, 1989, 16 (04) :609-613
[3]   MEASUREMENT OF DOSE IN THE BUILDUP REGION USING FIXED-SEPARATION PLANE-PARALLEL IONIZATION CHAMBERS [J].
GERBI, BJ ;
KHAN, FM .
MEDICAL PHYSICS, 1990, 17 (01) :17-26
[4]   EFFECT OF VARIOUS ATOMIC NUMBER ABSORBERS ON SKIN DOSE FOR 10-MEV X-RAYS [J].
KHAN, FM ;
MOORE, VC ;
LEVITT, SH .
RADIOLOGY, 1973, 109 (01) :209-212
[5]   USE OF ELECTRON FILTER TO REDUCE SKIN DOSE IN COBALT TELETHERAPY [J].
KHAN, FM .
AMERICAN JOURNAL OF ROENTGENOLOGY RADIUM THERAPY AND NUCLEAR MEDICINE, 1971, 111 (01) :180-&
[6]   DETERMINATION OF BUILDUP REGION OVER-RESPONSE CORRECTIONS FOR A MARKUS-TYPE CHAMBER [J].
MELLENBERG, DE .
MEDICAL PHYSICS, 1990, 17 (06) :1041-1044
[7]  
MONDALEK PM, 1987, MED DOSIMET, V7, P5
[9]   METHODS FOR ENHANCEMENT OF SKIN SPARING IN COBALT-60 TELETHERAPY [J].
SAYLOR, WL ;
QUILLIN, RM .
AMERICAN JOURNAL OF ROENTGENOLOGY RADIUM THERAPY AND NUCLEAR MEDICINE, 1971, 111 (01) :174-+