REDUCTION OF THE BREMSSTRAHLUNG COMPONENT OF CLINICAL ELECTRON-BEAMS - IMPLICATIONS FOR ELECTRON ARC THERAPY AND TOTAL SKIN ELECTRON-IRRADIATION

被引:12
作者
ELKHATIB, EE
SCRIMGER, J
MURRAY, B
机构
[1] Dept. of Phys., Cross Cancer Inst., Edmonton, Alta.
关键词
D O I
10.1088/0031-9155/36/1/010
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The dose due to Bremsstrahlung in stationary electron beams of nominal energies in the range 6-20 MeV is typically between 1-7% of the maximum dose and is usually not clinically significant. However, in treatments using rotational or multiple electron beams where the x-ray dose from several beams is added the x-ray dose will reach much higher proportions and will be of clinical significance. Moreover, this dose often is located in normal tissue beyond the target volume. Reduction of this x-ray dose is therefore desirable. In the present study a reduction of the x-ray component of electron beams produced by a Clinac 2100C accelerator by a change of the transmission ion chamber and scattering foils is reported. A reduction in Bremsstrahlung of up to 50% can be achieved.
引用
收藏
页码:111 / 118
页数:8
相关论文
共 14 条
[1]   METHOD FOR RAPID-DETERMINATION OF ENERGY OF ELECTRON-BEAMS FROM MEDICAL LINEAR ACCELERATORS [J].
BELL, JH ;
WAGGENER, RG .
RADIOLOGY, 1978, 126 (03) :787-793
[2]  
BRAHME A, 1982, P S ELECTRON DOSIMET, P103
[3]   A TECHNIQUE FOR TOTAL SKIN ELECTRON-IRRADIATION USING 6 LARGE FLATTENED ELECTRON-BEAMS - IMPLEMENTATION ON LINEAR ACCELERATORS HAVING DIFFERENT MODES OF ELECTRON-BEAM PRODUCTION [J].
ELKHATIB, EG ;
RYMEL, J ;
ALMOKHLEF, J .
BRITISH JOURNAL OF RADIOLOGY, 1989, 62 (740) :744-748
[4]  
Gur D, 1979, Med Phys, V6, P145, DOI 10.1118/1.594525
[5]   ADVANCES IN LINEAR-ACCELERATOR DESIGN FOR RADIOTHERAPY [J].
KARZMARK, CJ .
MEDICAL PHYSICS, 1984, 11 (02) :105-128
[6]   BREMSSTRAHLUNG DOSE TO PATIENTS IN ROTATIONAL ELECTRON THERAPY [J].
KASE, KR ;
BJARNGARD, BE .
RADIOLOGY, 1979, 133 (02) :531-532
[7]   ELECTRON ARC THERAPY - PHYSICAL MEASUREMENT AND TREATMENT PLANNING TECHNIQUES [J].
LEAVITT, DD ;
PEACOCK, LM ;
GIBBS, FA ;
STEWART, JR .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1985, 11 (05) :987-999
[8]   THE STUDY OF A SWEPT ELECTRON-BEAM IN ORDER TO APPLY BOAG THEORY FOR CALCULATION OF THE COLLECTION EFFICIENCY .1. BEAM AND SWEPT AREA CHARACTERISTICS [J].
MARINELLO, G ;
VALERO, M ;
DELPLANQUE, JM .
PHYSICS IN MEDICINE AND BIOLOGY, 1986, 31 (08) :859-868
[9]   RADIATION PARAMETERS OF 6 TO 20-MEV SCANNING ELECTRON-BEAMS FROM THE SATURNE LINEAR-ACCELERATOR [J].
PFALZNER, PM ;
CLARKE, HC .
MEDICAL PHYSICS, 1982, 9 (01) :117-120
[10]   ELECTRON DOSE-RATE AND PHOTON CONTAMINATION IN ELECTRON ARC THERAPY [J].
PLA, M ;
PODGORSAK, EB ;
PLA, C .
MEDICAL PHYSICS, 1989, 16 (05) :692-697