VERIFICATION OF REVISED WATER-VAPOR CORRECTION TO THE INDEX OF REFRACTION OF AIR

被引:19
作者
BEERS, J
DOIRON, T
机构
[1] Precision Engineering Division, National Institute of Standards and Technology, Gaithersburg
关键词
7;
D O I
10.1088/0026-1394/29/4/008
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The Edlen equation has become the standard method of relating the index of refraction of air to the air temperature, pressure and humidity. The recent suggestion for improvements in the humidity dependence by Birch and Downs has been verified at the 0,633 mum He-Ne laser wavelength using the line scale interferometer at the National Institute of Standards and Technology (NIST).
引用
收藏
页码:315 / 316
页数:2
相关论文
共 6 条
[1]  
Barrell H., 1939, PHIL T ROY SOC LOND, V238, P1
[2]  
BEERS JS, 1987, NBSIR873625 NBS
[3]   THE RESULTS OF A COMPARISON BETWEEN CALCULATED AND MEASURED VALUES OF THE REFRACTIVE-INDEX OF AIR [J].
BIRCH, KP ;
DOWNS, MJ .
JOURNAL OF PHYSICS E-SCIENTIFIC INSTRUMENTS, 1988, 21 (07) :694-695
[4]  
Edlen B., 1966, METROLOGIA, V2, P71, DOI [10.1088/0026-1394/2/2/002, DOI 10.1088/0026-1394/2/2/002]
[5]  
JONES FE, 1980, J RES NBS, V86, P27
[6]  
PENZES WB, 1990, SEP P S MEAS INSP IN, P41