A 3D MODEL OF NONUNIFORM ATTENUATION AND DETECTOR RESPONSE FOR EFFICIENT ITERATIVE RECONSTRUCTION IN SPECT

被引:31
作者
GILLAND, DR
JASZCZAK, RJ
WANG, H
TURKINGTON, TG
GREER, KL
COLEMAN, RE
机构
[1] Dept. of Radiol., Duke Univ. Med. Center, Durham, NC
关键词
D O I
10.1088/0031-9155/39/3/017
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A 3D physical model for iterative reconstruction in SPECT has been developed and applied to experimental data. The model incorporates non-uniform attenuation using reconstructed transmission cr data and distance-dependent detector response based on response function measurements over a range of distances from the detector. The 3D model has been implemented in a computationally efficient manner with practical memory requirements. The features of the model that provide efficiency are described including a new region-dependent reconstruction (RDR) technique. With RDR, filtered backprojection is used to reconstruct areas of the image of minimal clinical importance, and the result is used to supplement the iterative reconstruction of the clinically important areas of the image. The 3D model was incorporated into the maximum likelihood-expectation maximization (ML-EM) reconstruction algorithm and tested in three phantom studies-a point source, a uniform cylinder, and an anthropomorphic thorax-and a patient Tc-99(m) sestamibi study. Reconstructed images with the 3D method exhibited excellent noise and resolution characteristics. With the sestamibi data, the RDR technique produced essentially the conventional ML-EM estimate in the cardiac region with substantial time savings.
引用
收藏
页码:547 / 561
页数:15
相关论文
共 21 条
[1]  
BOWSHER J E, 1992, Journal of Nuclear Medicine, V33, P879
[3]   STOCHASTIC RELAXATION, GIBBS DISTRIBUTIONS, AND THE BAYESIAN RESTORATION OF IMAGES [J].
GEMAN, S ;
GEMAN, D .
IEEE TRANSACTIONS ON PATTERN ANALYSIS AND MACHINE INTELLIGENCE, 1984, 6 (06) :721-741
[4]  
GILLAND DR, 1991, J NUCL MED, V32, P527
[5]   QUANTITATIVE SPECT BRAIN IMAGING - EFFECTS OF ATTENUATION AND DETECTOR RESPONSE [J].
GILLAND, DR ;
JASZCZAK, RJ ;
BOWSHER, JE ;
TURKINGTON, TG ;
LIANG, Z ;
GREER, KL ;
COLEMAN, RE .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1993, 40 (03) :295-299
[6]   ALGEBRAIC RECONSTRUCTION TECHNIQUES (ART) FOR 3-DIMENSIONAL ELECTRON MICROSCOPY AND X-RAY PHOTOGRAPHY [J].
GORDON, R ;
BENDER, R ;
HERMAN, GT .
JOURNAL OF THEORETICAL BIOLOGY, 1970, 29 (03) :471-&
[7]   AN ATTENUATED PROJECTOR BACKPROJECTOR FOR ITERATIVE SPECT RECONSTRUCTION [J].
GULLBERG, GT ;
HUESMAN, RH ;
MALKO, JA ;
PELC, NJ ;
BUDINGER, TF .
PHYSICS IN MEDICINE AND BIOLOGY, 1985, 30 (08) :799-816
[8]  
HUDSON H M, 1992, Journal of Nuclear Medicine, V33, P960
[9]  
JASZCZAK RJ, 1993, J NUCL MED, V34, P1577
[10]  
JASZCZAK RJ, 1984, J NUCL MED, V25, P893