COMPARISON OF MATHEMATICAL AND MECHANICAL MODELS OF PRESSURE-CONTROLLED VENTILATION

被引:21
作者
BURKE, WC
CROOKE, PS
MARCY, TW
ADAMS, AB
MARINI, JJ
机构
[1] UNIV MINNESOTA,ST PAUL RAMSEY MED CTR,640 JACKSON ST,ST PAUL,MN 55101
[2] VANDERBILT UNIV,DEPT MATH,NASHVILLE,TN 37240
关键词
MATHEMATICAL MODELING; MECHANICAL VENTILATION; DYNAMIC HYPERINFLATION; EXPONENTIAL KINETICS;
D O I
10.1152/jappl.1993.74.2.922
中图分类号
Q4 [生理学];
学科分类号
071003 ;
摘要
Recent evidence that volume-cycled mechanical ventilation may itself produce lung injury has focused clinical attention on the pressure waveform applied to the respiratory system. There has been an increasing use of pressure-controlled ventilation (PCV), because it limits peak cycling pressure and provides a decelerating flow profile that may improve ps exchange. In this mode, however, the relationships are of machine adjustments to ventilation and alveolar pressure are not straightforward. Consequently, setting selection remains largely an empirical process. In previous work, we developed a biexponential model of PCV that provides a conceptual framework for understanding these interactions (J. Appl. Physiol. 67: 1081-1092, 1989). We tested the validity of this mathematical model in a single-compartment analogue of the respiratory system across wide ranges of clinician-set variables (frequency, duty cycle, applied pressure) and impedance conditions (inspiratory and expiratory resistance and system compliance). Our data confirm the quantitative validity of the proposed model when approximately rectilinear waves of pressure are applied and appropriate values for impedance are utilized. Despite a fixed-circuit configuration, however, resistance proved to be a function of each clinician-set variable, requiring remeasurement of system impedance as adjustments in these variables were made. With further modification, this model may provide a practical as well as a conceptual basis for understanding minute ventilation and alveolar pressure fluctuations during PCV in the clinical setting.
引用
收藏
页码:922 / 933
页数:12
相关论文
共 20 条
[1]   CARDIORESPIRATORY EFFECTS OF PRESSURE CONTROLLED VENTILATION IN SEVERE RESPIRATORY-FAILURE [J].
ABRAHAM, E ;
YOSHIHARA, G .
CHEST, 1990, 98 (06) :1445-1449
[2]  
ALSAADY N, 1985, INTENS CARE MED, V11, P68
[3]  
Baker A B, 1971, Int Anesthesiol Clin, V9, P39, DOI 10.1097/00004311-197100920-00004
[4]   ANALOG STUDY OF CONTROLLED VENTILATION [J].
BAKER, AB ;
HAHN, CEW .
RESPIRATION PHYSIOLOGY, 1974, 22 (03) :227-239
[5]  
BAUM M, 1980, ANAESTHESIST, V29, P592
[6]   INTRAPULMONARY GAS TRAPPING DURING MECHANICAL VENTILATION AT RAPID FREQUENCIES [J].
BERGMAN, NA .
ANESTHESIOLOGY, 1972, 37 (06) :626-&
[7]  
BERGMAN NA, 1984, ACTA ANAESTH SCAND, V28, P178
[8]  
BOROS SJ, 1984, PEDIATRICS, V74, P487
[9]  
BOYSEN P G, 1988, Respiratory Care, V33, P126
[10]   ELECTRICAL ANALOGUE OF LUNG [J].
CAMPBELL, D ;
BROWN, J .
BRITISH JOURNAL OF ANAESTHESIA, 1963, 35 (11) :684-&