INTRALUMINAL DEVICE PRESSURES IN 3-PIECE INFLATABLE PENILE PROSTHESES - THE PATHOPHYSIOLOGY OF MECHANICAL MALFUNCTION

被引:19
作者
PESCATORI, ES
GOLDSTEIN, I
机构
[1] Department of Urology, Boston University School of Medicine, Boston, MA
关键词
PENIS; IMPOTENCE; PENILE PROSTHESES;
D O I
10.1016/S0022-5347(17)36061-5
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
The role of intraluminal device pressures in the development of mechanical failures in 3-piece inflatable penile prostheses was investigated in a 2-part study. An in vitro study was performed in which mean intraluminal device pressures were recorded in Mentor IPP, Mentor Alpha-1, AMS 700 CX and AMS Ultrex devices. At maximum inflated volumes mean intraluminal device pressures exceeded 600 and 1,000 mm. Hg at rest, and 850 and 1,300 mm. Hg following external loading in Mentor and AMS products, respectively. The AMS 700 CX was associated with the highest mean intraluminal device pressure, with values approaching 1,400 mm. Hg following external loading. A 4-year clinical review at our institution was performed comparing the mechanical malfunctions in 51, 3-piece implants with connectors between the pump and cylinders (AMS 700 CX) to the mechanical malfunctions of 50, 3-piece implants with pre-connected tubing between the pump and cylinders (Mentor Alpha-1). Fluid leaks were identified in 16% of the former and 4% of the latter devices. It is proposed that the high intraluminal device pressures are an important factor in the pathophysiology of the mechanical malfunctions in 3-piece inflatable penile prostheses and that connectors are considered components at risk for pressure-associated fluid leaks. The ideal penile prosthesis should maintain the superior rigidity and cosmetic features of the 3-piece inflatable penile prosthesis. However, this device should be pre-connected and designed to mimic more closely the anatomy of the human penis to achieve functional erections at lower intraluminal device pressure values.
引用
收藏
页码:295 / 300
页数:6
相关论文
共 21 条
[1]  
BRIGHT TC, 1978, CAMPBELLS UROLOGY, V1, P906
[2]   COMPLICATIONS ASSOCIATED WITH PENILE IMPLANTS USED TO TREAT IMPOTENCE [J].
EARLE, CM ;
WATTERS, GR ;
TULLOCH, AGS ;
WISNIEWSKI, ZS ;
LORD, DJ ;
KEOGH, EJ .
AUSTRALIAN AND NEW ZEALAND JOURNAL OF SURGERY, 1989, 59 (12) :959-962
[3]   CHARACTERIZATION OF PENILE ERECTILE STATES USING EXTERNAL COMPUTER-BASED MONITORING [J].
FROHRIB, DA ;
GOLDSTEIN, I ;
PAYTON, TR ;
PADMANATHAN, H ;
KRANE, RJ .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1987, 109 (02) :110-114
[4]   THE INFLATABLE PENILE PROSTHESIS - CLINICAL-EXPERIENCE WITH A NEW CONTROLLED EXPANSION CYLINDER [J].
FURLOW, WL ;
MOTLEY, RC .
JOURNAL OF UROLOGY, 1988, 139 (05) :945-946
[5]   THE FIBROUS SKELETON OF THE CORPORA CAVERNOSA AND ITS PROBABLE FUNCTION IN THE MECHANISM OF ERECTION [J].
GOLDSTEIN, AMB ;
MEEHAN, JP ;
MORROW, JW ;
BUCKLEY, PA ;
ROGERS, FA .
BRITISH JOURNAL OF UROLOGY, 1985, 57 (05) :574-578
[6]   5-YEAR FOLLOW-UP OF THE SCOTT INFLATABLE PENILE PROSTHESIS AND COMPARISON WITH SEMIRIGID PENILE PROSTHESIS [J].
KABALIN, JN ;
KESSLER, R .
JOURNAL OF UROLOGY, 1988, 140 (06) :1428-1430
[7]   COMPLICATIONS OF INFLATABLE PENILE PROSTHESES [J].
KESSLER, R .
UROLOGY, 1981, 18 (05) :470-472
[8]   PATIENT AND PARTNER SATISFACTION WITH THE AMS-700 PENILE PROSTHESIS [J].
MCLAREN, RH ;
BARRETT, DM .
JOURNAL OF UROLOGY, 1992, 147 (01) :62-65
[9]   CLINICAL-EXPERIENCE WITH THE MENTOR INFLATABLE PENILE PROSTHESIS IN 301 PATIENTS [J].
MERRILL, DC .
JOURNAL OF UROLOGY, 1988, 140 (06) :1424-1427
[10]   CLINICAL-EXPERIENCE WITH SCOTT INFLATABLE PENILE PROSTHESIS IN 150 PATIENTS [J].
MERRILL, DC .
UROLOGY, 1983, 22 (04) :371-375