IMPROVING THE PERFORMANCE OF ULTRAFAST MICROCHANNEL-PLATE PHOTOMULTIPLIERS IN TIME-CORRELATED PHOTON-COUNTING BY PULSE PRESHAPING

被引:2
作者
COVA, S [1 ]
RIPAMONTI, G [1 ]
机构
[1] CNR,CTR ELETTR QUANTIST & STRUMENTAZ,I-20133 MILAN,ITALY
关键词
D O I
10.1063/1.1141426
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
With ultrafast microchannel-plate (MCP) photomultipliers that produce subnanosecond pulses, the commercially available circuit modules do not provide a correct constant-fraction trigger (CFT) operation. This has a detrimental effect on the resolution obtained in time-correlated photon counting experiments. In order to circumvent this CFT limitation and better exploit the ultrafast MCP performance, a simple pulse-shaping filter has been devised and experimented with. The filter can be interposed between the MCP output and the input of any commercial CFT module, without modifying the latter. The benefit of the filter is confirmed by the experimental results. With MCP samples that produced awkward resolution curves with satellite peaks and other irregular features, the introduction of the filter regularized and narrowed the curve. Also with a setup having fairly good performance, the introduction of the filter was advantageous: the full width at half maximum (FWHM) resolution improved from 75 to 55 ps.
引用
收藏
页码:1072 / 1075
页数:4
相关论文
共 8 条
[1]   MINIMIZATION OF DISCRIMINATION JITTER IN TIME-CORRELATED SINGLE PHOTON-COUNTING [J].
ANFINRUD, PA ;
STRUVE, WS .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1989, 60 (04) :800-800
[3]   SYNTHESIS OF AN OPTIMUM FILTER FOR TIMING SCINTILLATION PULSES [J].
GATTI, E ;
SVELTO, V .
NUCLEAR INSTRUMENTS & METHODS, 1966, 39 (02) :309-&
[4]   DESIGN OF CONSTANT FRACTION OF PULSE HEIGHT TRIGGER FOR OPTIMUM TIME RESOLUTION [J].
GEDCKE, DA ;
MCDONALD, WJ .
NUCLEAR INSTRUMENTS & METHODS, 1968, 58 (02) :253-+
[5]   A CONSTANT FRACTION OF PULSE HEIGHT TRIGGER FOR OPTIMUM TIME RESOLUTION [J].
GEDCKE, DA ;
MCDONALD, WJ .
NUCLEAR INSTRUMENTS & METHODS, 1967, 55 (02) :377-+
[6]   ULTRAFAST MICROCHANNEL PLATE PHOTOMULTIPLIERS [J].
KUME, H ;
KOYAMA, K ;
NAKATSUGAWA, K ;
SUZUKI, S ;
FATLOWITZ, D .
APPLIED OPTICS, 1988, 27 (06) :1170-1178
[7]  
NICHOLSON PW, 1974, NUCL ELECTRON, P259
[8]  
OCONNOR DV, 1984, TIME CORRELATED SING, P135