SCANNING PROCEDURES FOR CALIBRATION OF RADIATION DETECTORS

被引:4
作者
BIRATTARI, C [1 ]
MANFREDI, G [1 ]
SILARI, M [1 ]
机构
[1] CNR,IST TECNOL BIOMED AVANZATE,I-20131 MILAN,ITALY
关键词
D O I
10.1063/1.1145370
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Three different scanning procedures that can be adopted to reproduce a uniform irradiation by a radiation field over a large detector, when the beam size is less than the monitor dimension, are discussed. These procedures are spot scanning, in which the scan is performed over a mesh of points; linear scanning, in which the instrument is moved continuously in front of the beam in the two transverse directions; and cross scanning, in which the scan is obtained according to a law yielding a ''pseudo-Lissajous'' figure. An instrument which has been designed and built for the calibration of a neutron rem counter is described. (C) 1995 American Institute of Physics.