ON LINE SPECIATION OF SE(VI), SE(IV), AND TRIMETHYLSELENIUM BY HPLC MICROWAVE-OVEN HYDRIDE GENERATION ATOMIC-ABSORPTION SPECTROMETRY

被引:72
作者
COBOFERNANDEZ, MG
PALACIOS, MA
CHAKRABORTI, D
QUEVAUILLER, P
CAMARA, C
机构
[1] UNIV COMPLUTENSE MADRID,FAC CIENCIAS QUIM,DEPT QUIM ANALIT,E-28040 MADRID,SPAIN
[2] JADAVPUR UNIV,SCH ENVIRONM STUDIES,CALCUTTA 700032,W BENGAL,INDIA
[3] COMMUNITY BUR REFERENCE,B-1049 BRUSSELS,BELGIUM
来源
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY | 1995年 / 351卷 / 4-5期
关键词
D O I
10.1007/BF00322915
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An on-line system is proposed consisting of an anion-exchange chromatographic column, microwave-induced thermooxidation of trimethylselenium in the presence of persulphate, and microwave-induced thermoreduction of Se(VI) to Se(IV) in HCl medium, followed by hydride generation and atomic absorption for the determination of trimethylselenium (TMeSe), Se(IV) and Se(VI). Trimethylselenium is eluted in the dead volume of an anion-exchange column (Hamilton PRP-X-100), before elution of Se(IV) and Se(VI). Optimum chromatographic conditions have been obtained using 100 mmolL(-1) phosphate buffer (pH = 6.8) H2PO4-/HPO42- as the mobile phase. Recoveries were around 100%, absolute detection limits were 1.1, 1.4 and 2.2 ng for TMeSe, Se(IV) and Se(VI), respectively. Precision was lower than 10% in all cases. The method has been applied to tap water.
引用
收藏
页码:438 / 442
页数:5
相关论文