The direct determination of As in solid samples of plant origin using electrothermal vaporization inductively coupled plasma mass spectrometry (ETV-ICP-MS) is reported. The first phase of the work consisted of an optimization of the carrier gas bow rate and the heating cycle of the modified graphite furnace lashing and vaporization temperatures). A systematic study of the signal profiles (signal intensity as a function of time) showed that As coming from the solid samples and As coming from Liquid standards added to the solid material, only showed an analogous behaviour on condition that these liquid spikes were previously dried before the solid sample was introduced into the sample boat. The potential of the technique was assessed by establishing figures of merit and analysing some standard reference materials for As, An absolute limit of detection of approximately 1 pg was established for As, corresponding to a relative limit of detection of approximately 1 ng g(-1) for a typical sample mass of 1 mg. Results obtained for the As content in the samples analysed were not deteriorated by the presence of Cl, as even on addition of amounts of Cl exceeding the Cl content of the samples no (ArCl+)-Ar-40-Cl-35 interference on the As-75(+) signal could be established. For the determination of As in these reference materials of plant origin, several methods of calibration were investigated, including external calibration using both liquid and solid standards and single standard addition. The use of an internal standard (Sb) was studied and its requirement was assessed as being from advisable to imperative, depending on the calibration method used. Although several methods offered possibilities for accurate determination (mean deviation between ETV-ICP-MS results and certified values < 10%), single standard addition was assessed to be the most practicable and straightforward method.
引用
收藏
页码:81 / 87
页数:7
相关论文
共 45 条
[1]   LASER ABLATION OF SOLIDS FOR ELEMENTAL ANALYSIS BY INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
ARROWSMITH, P .
ANALYTICAL CHEMISTRY, 1987, 59 (10) :1437-1444
[2]   INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION-SPECTROMETRY OF POWDERED SAMPLES USING THE MINIATURE CUP TECHNIQUE [J].
ATSUYA, I ;
ITOH, T ;
KUROTAKI, T .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1991, 46 (01) :103-108
[4]   PROBE DESIGN FOR THE DIRECT INSERTION OF SOLID SAMPLES IN THE INDUCTIVELY COUPLED PLASMA FOR ANALYSIS BY ATOMIC EMISSION AND MASS-SPECTROMETRY [J].
BLAIN, L ;
SALIN, ED ;
BOOMER, DW .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1989, 4 (08) :721-725
[5]   APPLICATION OF A WIRE LOOP DIRECT SAMPLE INSERTION DEVICE FOR INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
BOOMER, DW ;
POWELL, M ;
SING, RLA ;
SALIN, ED .
ANALYTICAL CHEMISTRY, 1986, 58 (04) :975-976
[6]   USE OF THE SIMPLIFIED GENERALIZED STANDARD ADDITIONS METHOD FOR CALIBRATION IN SOLID SAMPLING ELECTROTHERMAL VAPORIZATION INDUCTIVELY-COUPLED PLASMA-ATOMIC EMISSION-SPECTROMETRY [J].
BOONEN, S ;
VERREPT, P ;
MOENS, LJ ;
DAMS, RFJ .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1993, 8 (05) :711-714
[7]   A STUDY OF SOME TECHNIQUES FOR DIRECT SOLIDS SAMPLING IN PLASMA SPECTROMETRY [J].
BROEKAERT, JAC ;
LEIS, F ;
RAEYMAEKERS, B ;
ZARAY, G .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1988, 43 (4-5) :339-353
[8]   SAMPLE INTRODUCTION TECHNIQUES FOR ATOMIC SPECTROSCOPY [J].
BROWNER, RF ;
BOORN, AW .
ANALYTICAL CHEMISTRY, 1984, 56 (07) :A875-+
[9]  
BROWNER RF, 1984, ANAL CHEM, V56, pA787
[10]   ELECTROTHERMAL VAPORIZATION FOR SAMPLE INTRODUCTION IN PLASMA SOURCE SPECTROMETRY [J].
CAREY, JM ;
CARUSO, JA .
CRITICAL REVIEWS IN ANALYTICAL CHEMISTRY, 1992, 23 (05) :397-439