DIGITAL FILTERING OF ONLINE EVOKED-POTENTIALS

被引:6
作者
ACKMANN, JJ [1 ]
ELKO, PP [1 ]
WU, SJ [1 ]
机构
[1] MARQUETTE UNIV,MILWAUKEE,WI 53233
来源
INTERNATIONAL JOURNAL OF BIO-MEDICAL COMPUTING | 1979年 / 10卷 / 04期
关键词
D O I
10.1016/0020-7101(79)90002-3
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
Evoked potentials recorded automatically at frequent intervals are a useful adjunct for monitoring head injury patients; however, unaveraged residual noise due to patient movement and synchronization of the stimulus to harmonics of the line frequency is sometimes present. A frequency analysis was performed on 23 records with varying degrees of unaveraged noise and the results were used to design a digital filter. The frequency content of the records analyzed was largely contained in a band from D.C. to 200 Hz. A zero phase shift lowpass digital filter with a cutoff frequency of 200 Hz eliminated the majority of noise. Despite the low frequency content, a wide-band amplifier is still required to avoid waveform distortion, particularly latencies. Digital filters can then be applied without causing distortion. © 1979.
引用
收藏
页码:291 / 303
页数:13
相关论文
共 7 条
  • [1] ACKMANN JJ, 1975, 28TH P ANN C ENGNG M
  • [2] Bendat J. S., 1971, RANDOM DATA ANAL MEA
  • [3] SYSTEM BANDPASS REQUIRED TO AVOID DISTORTION OF EARLY COMPONENTS WHEN AVERAGING SOMATOSENSORY EVOKED-POTENTIALS
    DESMEDT, JE
    BRUNKO, E
    DEBECKER, J
    CARMELIET, J
    [J]. ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY, 1974, 37 (04): : 407 - 410
  • [4] EVOKED SOMATOSENSORY POTENTIALS IN MAN
    LARSON, SJ
    SANCES, A
    CHRISTENSON, PC
    [J]. ARCHIVES OF NEUROLOGY, 1966, 15 (01) : 88 - +
  • [5] LARSON SJ, 1973, SURGERY, V74, P34
  • [6] RABINER LR, 1975, THEORY APPLICATION D
  • [7] STEARNS SD, 1975, DIGITAL SIGNAL ANAL