POLYETHYLENE AND METAL DEBRIS GENERATED BY NON-ARTICULATING SURFACES OF MODULAR ACETABULAR COMPONENTS

被引:144
作者
HUK, OL
MANJULA, B
BETTS, F
RIMNAC, CM
LIEBERMAN, JR
HUO, MH
SALVATI, EA
机构
[1] HOP HOTEL DIEU,MONTREAL H2W 1T8,PQ,CANADA
[2] NEW YORK HOSP,HOSP SPECIAL SURG,NEW YORK,NY 10021
[3] CORNELL UNIV,COLL MED,NEW YORK,NY 10021
[4] UNIV CALIF LOS ANGELES,CTR HLTH SCI,SCH MED,DEPT ORTHOPAED SURG,LOS ANGELES,CA 90024
[5] WATERBURY HOSP & HLTH CTR,CTR JOINT REPLACEMENT,WATERBURY,CT
来源
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME | 1994年 / 76B卷 / 04期
关键词
D O I
10.1302/0301-620X.76B4.8027142
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
We report a prospective study of the liner-metal interfaces of modular uncemented acetabular components as sources of debris. We collected the pseudomembrane from the screw-cup junction and the empty screw holes of the metal backing of 19 acetabula after an average implantation of 22 months. Associated osteolytic lesions were separately collected in two cases. The back surfaces of the liners and the screws were examined for damage, and some liners were scanned by electron microscopy. The tissues were studied histologically and by atomic absorption spectrophotometry to measure titanium content. The pseudomembrane from the screw-cup junction contained polyethylene debris in seven specimens and metal debris in ten. The material from empty screw holes was necrotic tissue or dense fibroconnective tissue with a proliferative histiocytic infiltrate and foreign-body giant-cell reaction. It contained polyethylene debris in 14 cases and metal in five. The two acetabular osteolytic lesions also showed a foreign-body giant-cell reaction to particulate debris. The average titanium levels in pseudomembranes from the screw-cup junction and the empty screw holes were 959 mu g/g (48 to 11 900) and 74 mu g/g (0.72 to 331) respectively. The tissue from the two lytic lesions showed average titanium levels of 139 and 147 mu g/g respectively. The back surfaces of the PE liners showed surface deformation, burnishing, and embedded metal debris. All 30 retrieved screws demonstrated fretting at the base of the head and on the proximal shaft. Non-articular modular junctions create new interfaces for the generation of particulate debris, which may cause granulomatous reaction.
引用
收藏
页码:568 / 574
页数:7
相关论文
共 27 条
[1]   METALLIC WEAR IN FAILED TITANIUM-ALLOY TOTAL HIP REPLACEMENTS - A HISTOLOGICAL AND QUANTITATIVE-ANALYSIS [J].
AGINS, HJ ;
ALCOCK, NW ;
BANSAL, M ;
SALVATI, EA ;
WILSON, PD ;
PELLICCI, PM ;
BULLOUGH, PG .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1988, 70A (03) :347-356
[2]  
[Anonymous], 1992, SOCIOLOGICA, V7
[3]  
Bartel D L, 1983, Hip, P229
[5]   DISSOCIATION OF ACETABULAR COMPONENTS AFTER TOTAL HIP-ARTHROPLASTY - REPORT OF 4 CASES [J].
BRIEN, WW ;
SALVATI, EA ;
WRIGHT, TM ;
NELSON, CL ;
HUNGERFORD, DS ;
GILLIAM, DL .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1990, 72A (10) :1548-1550
[6]  
Bueche M J, 1989, J Arthroplasty, V4, P39, DOI 10.1016/S0883-5403(89)80052-X
[7]   TOTAL HIP REPLACEMENT FAILURES - HISTOLOGICAL EVALUATION [J].
CHAROSKY, CB ;
BULLOUGH, PG ;
WILSON, PD .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1973, A 55 (01) :49-58
[8]  
Collier J. P., 1992, CLIN ORTHOPAEDICS, V285, P129
[9]   THE INITIAL STABILITY OF UNCEMENTED ACETABULAR COMPONENTS [J].
CURTIS, MJ ;
JINNAH, RH ;
WILSON, VD ;
HUNGERFORD, DS .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1992, 74 (03) :372-376
[10]   CHARACTERISTICS OF ACETABULAR CUPS WORN IN HUMAN-BODY [J].
DOWLING, JM ;
ATKINSON, JR ;
DOWSON, D ;
CHARNLEY, J .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1978, 60 (03) :375-382